CAUTION
This product is intended for adult use only.
This product is intended for non-invasive measuring and monitoring of arterial blood
pressure.
It is not intended for the use on extremities other than the arm or for functions other than
obtaining a blood pressure measurement.
Do not confuse self-monitoring with self-diagnosis. This product allows you to monitor your
blood pressure. Do not begin or end medical treatment without asking a physician for
treatment advice.
If you are taking medication, consult your physician to determine the most appropriate time
to measure your blood pressure. Never change a prescribed medication without consulting
your Physician.
When the product was used to measure patients who have common arrhythmias such as
atrial or ventricular premature beats or atrial fibrillation, the best result may occur with
deviation. Please consult your physician about the result.
If the cuff pressure exceeds 40 kPa (300 mmHg), the product will automatically deflate.
Should the cuff not deflate when pressures exceeds 40 kPa (300 mmHg), detach the cuff
from the arm and press the START/STOP button to stop inflation.
The product is not AP/APG equipment and not suitable for use in the presence of a
flammable anesthetic mixture with air of with oxygen or nitrous oxide.
The operator shall not touch output of batteries and the patient simultaneously.
To avoid errors in measurement, please avoid the condition of strong electromagnetic field
radiated interference signal or electrical fast transient/burst signal.
Check that the product's functions carefully and ensure that it is in proper working condition
before being used.
This product is contraindicated for any female who may be suspected of, or is pregnant.
Besides providing inaccurate readings, the effects of this product on the fetus are
unknown.
Manufacturer will make available on request circuit diagrams, component parts list etc.
This product is not suitable for continuous monitoring during medical emergencies or
operations. Otherwise, the patient’s arm and fingers will become anaesthetic, swollen and
even purple due to a lack of blood.
Read the instruction manual and use the product properly.
Otherwise, the performance and lifetime of the product will be impacted and reduced.
During use, the patient will be in contact with the cuff. The materials of the cuff have been
tested and found to comply with requirements of ISO 10993-5:2009 and ISO
10993-10:2010. It will not cause any potential sensization or irritation reaction.
Please use ACCESSORIES and detachable parts specified/ authorised by
MANUFACTURER. Otherwise, it may cause damage to the product or danger to the user /
patients.
The product doesn’t need to be calibrated within the two years of reliable service.
Please dispose of ACCESSORIES, detachable parts, and the ME EQUIPMENT according
to the local guidelines.
Please use the soft cloth to clean the whole product. Don’t use any abrasive or volatile
cleaners.
3
INTRODUCTION