Alarms
Inoperable Alarms
Inoperable alarms (table 8) are medium priority alarms. To correct the condition, perform these
steps:
1. Turn the monitor off and then back on again to remove the error message.
2. If the error persists, note the error code and contact Nonin Technical Service at (800) 356-
8874 (USA and Canada), +1 (763) 553-9968, or +31 (0)13 - 79 99 040 (Europe).
Low Priority Alarms
Low priority alarms do not have an audible alarm.
Table 8. Inoperable Alarms
Message/Error Code
Probable Cause
NO CAPNO E04
No communication from capnography unit.
SPEAKER FAULT E05
Speaker is not working properly.
SYSTEM FAILURE E06
General system error.
BUTTON FAILURE E07
Audio Pause button is damaged.
MEMORY CORRUPT E08
Error detected in memory structure.
Table 9. Low Priority Alarm
Alarm
Visual Indicator
Incompatible USB flash drive has been connected
to the monitor.
“Incompatible USB” pop-up window
displays.