17
Display Screens
Operating Screen
The main operating screen (figure 4) displays parameters, graphs, and other information.
NOTE:
If a pop-up window displays while monitoring, the icons and alarm limit up/down
arrows cannot be used until the pop-up window is closed.
Figure 4. Operating Screen
Table 4. Operating Screen – Display Descriptions
No.
Symbol/Name
Description
1
Alarm Limits
The high alarm limit is always located above the displayed
value, and the low alarm limit is always located below the
displayed value.
When the parameter readings fall between the low and high
settings, they are treated as normal values. Values outside
these limits activate both audible and visual alarms. The limit
that triggered the alarm flashes on the display.
See
for more information.
2
EtCO
2
mmHg
or
EtCO
2
kPa
Displays the volume of end tidal CO
2
in exhaled air. EtCO
2
is
shown as mmHg or kPa.
2
8
9
4
6
10
7
5
1
3
11