16
Using the 2500A VET Pulse Oximeter
Model 2000SA Small Animal Flex Sensor Application
The small animal flex sensor is suitable for use on the toe of a dog or the base of the tail on very
small animals such as rats.
Shave the site completely before sensor application. Position the light emitter and light detector of
the sensor so that the sensor light is directed through the tail or toe.
NOTE:
Mispositioning might allow sensor light to bypass the toe or tail resulting in SpO
2
measurement error. Secure the sensor with tape, ensuring that the tape does not restrict
perfusion.
Model 2000T Transflectance Sensor Application
The transflectance sensor is designed for placement on the underside, base of the tail (near the anal
orifice) of a dog.
Shave the site completely before sensor application. Position the light emitter and light detector of
the sensor against the underside of the tail at the base near the anal orifice. Secure the sensor with
tape, ensuring that the tape does not restrict perfusion.