Table 2: Electromagnetic Immunity
Manufacturer’s Declaration
Refer to the following tables for specific information regarding this device’s compliance to IEC 60601-1-2.
Table 1: Electromagnetic Emissions
Emissions Test
Compliance
Electromagnetic Environment—Guidance
This device is intended for use in the electromagnetic environment specified below.
The customer and/or user of this device should ensure that it is used in such an environment.
RF Emissions
CISPR 11
Group 1
This device uses RF energy only for its internal function. Therefore, its RF emissions are
very low and are not likely to cause any interference in nearby electronic equipment.
RF Emissions
CISPR 11
Class B
This device is suitable for use in all establishments, including domestic and those directly
connected to the public low-voltage power supply network that supplies buildings used for
domestic purposes.
Harmonic Emissions
IEC 61000-3-3
N/A
Voltage Fluctuations/Flicker Emissions
IEC 61000-3-3
N/A
Immunity Test
IEC 60601 Test Level
Compliance
Level
Electromagnetic
Environment—Guidance
This device is intended for use in the electromagnetic environment specified below.
The customer and/or user of this device should ensure that it is used in such an environment.
Electrostatic Discharge (ESD)
IEC 61000-4-2
±6 kV contact
±8 kV air
±6 kV contact
±8 kV air
Floors should be wood, concrete, or ceramic tile. If
floors are covered with synthetic material, relative
humidity should be at least 30%.
Electrical Fast Transient/Burst
IEC 61000-4-4
±2 kV for power supply lines
±1 kV for input/output lines
N/A
Mains power quality should be that of a typical
commercial or hospital environment.
Surge
IEC 61000-4-5
±1 kV differential mode
±2 kV common mode
N/A
Mains power quality should be that of a typical
commercial or hospital environment.
Voltage dips, short interruptions,
and voltage variations on power
supply input lines
IEC 61000-4-11
±5% U
T
(>95% dip in U
T
) for 0.5 cycle
±40% U
T
(60% dip in U
T
) for 5 cycles
±70% U
T
(30% dip in U
T
) for 25 cycles
<5% U
T
(>95% dip in U
T
) for 5 sec.
N/A
Mains power quality should be that of a typical
commercial or hospital environment.
Power Frequency
(50/60 Hz) Magnetic Field
IEC 61000-4-8
3 A/m
3 A/m
Power frequency magnetic fields should be at levels
characteristic of a typical location in a typical
commercial or hospital environment.
NOTE
: U
T
is the AC mains voltage before application of the test level.