Nonin 9571 Go2 LED User Manual Download Page 4

3

4

ENGLISH

Battery

Indicates separate collection for electrical and electronic equipment (WEEE).

IP33

Protected against spraying water and against access to hazardous parts  
with a tool, per IEC 60529.

SN

Serial Number

EC REP

Authorized Representative in the European Community

+70°C

 -30°C

Temperature Limitation for storage/shipping

%SpO

2

Percent Oxygen Saturation Display

Pulse Rate Display

Intended Use

GO

2

 

LED 

Model 9571 is intended to measure blood oxygen saturation (%SpO

2

) (the 

amount of oxygen in your blood) and pulse rate      of both adults and children. It is de-
signed for fingers (not the thumb) between 0.3 and 1.0 inch (0.8 – 2.5cm) thick. The 

GO

2

 

LED

 Model 9571 is intended for low perfusion conditions as defined in the specifications 

in this manual. Limit finger movement as much as possible when using this device. 

Con-

tact your licensed health care professional for your expected oxygen saturation level (to 
compare with your readings). The 

GO

2

 

LED 

Model 9571 is intended for Home Health 

Care Only.

Contraindications

  • 

Do not use the 

GO

2

 

LED 

in a magnetic resonance (MR) environment. 

Warnings

  • 

Keep the oximeter away from young children. Small items such as the battery door,  

    battery, and lanyard are choking hazards.

  • 

Certain activities may pose a risk of injury, including strangulation, if the lanyard    

    should become wrapped around your neck. Use the lanyard with caution.

7366-201-01 GO2 LED User Guide1_ENG.indd   3-4

3/12/2010   10:32:00 AM

Summary of Contents for 9571 Go2 LED

Page 1: ...9571 MODEL User Guide GO2 LED Users Guide1 Cover nospinelines indd 2 10 12 2009 3 25 41 PM...

Page 2: ...ger 10 Activating the Device and Verifying Operation 12 Reading Your Results 13 Care and Maintenance 15 Specifications 20 Equipment Response Time 23 Testing Summary 24 SpO2 Accuracy Testing 24 Pulse R...

Page 3: ...LED Model 9571 One AAA Alkaline Battery User Guide Quick Guide Back Cover Symbols The following symbols are associated with your GO2 LED Model 9571 Symbol Definition of Symbol Consult Instructions fo...

Page 4: ...inch 0 8 2 5cm thick The GO2 LED Model 9571 is intended for low perfusion conditions as defined in the specifications in this manual Limit finger movement as much as possible when using this device C...

Page 5: ...lowing conditions may reduce the performance of the GO2 LED flickering or very bright light weak pulse quality low perfusion low hemoglobin arterial catheters nail polish and or artificial nails and a...

Page 6: ...y of this pulse oximeter Follow local disposal and recycling laws for the GO2 LED and its components including the battery The GO2 LED is a precision electronic instrument and must be repaired by Noni...

Page 7: ...e will be stored for more than 30 days Applying the GO2 LED Model 9571 to Your Finger Hold the GO2 LED with the display facing toward you slide your finger into the opening at the bottom of the device...

Page 8: ...vides the correct pressure additional pressure may cause inaccurate readings Activating the Device and Verifying Operation The GO2 LED automatically turns on when a finger is inserted When a finger is...

Page 9: ...r reposition to another finger After the startup sequence the GO2 LED begins sensing the pulse indicated by the blinking pulse quality indicator Allow the device to stabilize observing about 4 seconds...

Page 10: ...re frequently if handled by multiple users Caution Do not place the GO2 LED in liquid or clean it with agents containing ammo nium chloride isopropyl alcohol or products that are not listed in this Us...

Page 11: ...low device to sit at room temperature for at least 30 minutes After unit has been dropped or shocked battery may need to be removed and reinstalled If the problem persists contact Nonin Technical Serv...

Page 12: ...ecifications Oxygen Saturation Display Range 0 to 100 Pulse Rate Display Range 18 to 321 beats per minute BPM Oxygen Saturation Declared Accuracy Range Arms 70 to 100 SpO2 2 digits Low Perfusion Oxyge...

Page 13: ...ion Type BF Applied Part Enclosure Degree of Ingress Protection IP33 Mode of Operation Continuous This equipment complies with International Standard IEC 60601 1 2 2004 for electromagnetic compatibili...

Page 14: ...cular requirements for the basic safety and essential performance of pulse oximeter equipment for medical use Pulse Rate Motion Testing This test measures pulse rate oximeter accuracy with motion arti...

Page 15: ...s GO2 LED Model 9571 is a precision electronic instrument and must be repaired by Nonin Technical Service Any sign or evidence of opening the GO2 LED Model 9571 field service by non Nonin personnel ta...

Page 16: ...if you have to contact Nonin with technical service issues or if you have any questions regarding the use or performance of your pulse oximeter Nonin s technical service department can be reached at...

Page 17: ...www go2nonin com 70 C 30 C 0123 Warning Keep away from small children this device contains small parts that may pose a choking hazard GO2 LED Users Guide1 Cover nospinelines indd 1 10 12 2009 3 25 38...

Reviews: