![Nonin 2500 PalmSAT Operator'S Manual Download Page 32](http://html1.mh-extra.com/html/nonin/2500-palmsat/2500-palmsat_operators-manual_1700051032.webp)
26
Technical Information
Table 6: Guidance and Manufacturer’s Declaration—Electromagnetic Immunity
Immunity Test
IEC 60601
Test Level
Compliance
Level
Electromagnetic Environment—
Guidance
This device is intended for use in the electromagnetic environment specified below. The user of this
device should ensure that it is used in such an environment.
Portable and mobile RF communications equipment should be used no closer to any part of the
device, including cables, than the recommended separation distance calculated from the equation
applicable to the frequency of the transmitter.
Recommended Separation Distance
Conducted RF
IEC 61000-4-6
3 V
rms
150 kHz to 80
MHz
3 V
Radiated RF
IEC 61000-4-3
3 V/m
80 MHz to 2.5
GHz
3 V/m
Radiated RF per ISO
9919 clause 36 and
ISO 80601-2-61
clause 202.6.2.3
20 V/m
80 MHz to 2.5
GHz
20 V/m
where
P
is the maximum output power
rating of the transmitter in watts (W)
according to the transmitter manufacturer
and
d
is the recommended separation
distance in meters (m).
Field strengths from fixed RF transmitters,
as determined by an electromagnetic site
survey
a
, should be less than the
compliance level in each frequency range.
b
Interference may occur in the vicinity of
equipment marked with the following
symbol:
a. Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios,
amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the
electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the mea-
sured field strength in the location in which the device is used exceeds the applicable RF compliance level above, the device
should be observed to verify normal operation. If abnormal performance is observed, additional measures may be neces-
sary, such as reorienting or relocating the device.
b. Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
NOTES:
• At 80 MHz and 800 MHz, the higher frequency range applies.
• These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects, and people.
d
1.17 P
=
d
1.17 P
=
d
2.33 P
=