Attributes
PROFILE A 15-100% On/Off 199°F (93°C) 133°F (56°C) 30 mins. F° 0 “Sea Level”
PROFILE B 15-100% On/Off 207°F (97°C) 131°F (55°C) 30 mins. F° 0 “Sea Level”
PROFILE C 15-100% On/Off 204°F (96°C) 134°F (57°C) 20 mins. F° 0 “Sea Level”
PROFILE D
15-100% On/Off 207°F (97°C) 131°F (55°C) 15 mins. F° 0 “Sea Level
Power Supply Interruption
In the event of a power interruption
(example: electrical outage due to weather),
the computer-controller will remember the
heating cycle and will continue with the
same cycle if the power supply comes
back online within 15 minutes. If the
colloid material is not solid, start a normal
cook-down cycle by turning the unit on and
pressing the ENTER button. If the power
supply is off for more than 15 minutes and
the colloid material is completely solid,
select the cold start cook-down cycle by
following these steps: With the power on,
press the SET button then the “ ” button
then the SET button twice (display will flash
“DONE”), then press ENTER twice.
VI. Settings
The factory default profile “A” is set-up
for Nobilium, Ticonium or Niranium colloid
products used at sea level. If the duplicator
will not be used at sea-level, the altitude
setting must be adjusted. Failure to set
the Altitude could cause the duplicator to
deliver poor quality colloid.
Select Pre-Programmed Setting
If not using Nobilium, Ticonium or Niranium
colloid products, review the requirements
of the duplicating material to be used and
match them to one of the pre-programmed
profile settings below. To select a specific
profile, the duplicator must be in “Standby
Mode” (power on but no heating lights illu-
minated). Press the SET button twice. The
display will show the first option on the set-
tings menu which is the present profile.
Press the “^” (UP) button to find the profile
to be changed. Press the SET button again
to select the profile. Press the ENTER button
to exit the settings menu.
Develop Custom Settings
If none of the pre-programmed settings
match the requirements of the duplicating
material, then develop a custom profile.
To develop custom profile settings, the
duplicator must be in “Standby Mode”
(power on but no heating lights illuminated).
Press the SET button once. Then press the
“^” button to find the attribute to be
changed. When the desired attribute is
displayed, press the SET button to select it.
Then press the “^” or “ ” button to change
the setting. Press the SET button to confirm
the change. Press the ENTER button to exit
the settings menu. The attribute setting
options are presented below along with their
maximum and minimum values.
To reset the duplicator to factory defaults,
scroll through the menu options until the
display reads “SET DEF.” Press the SET
button to return all settings to factory default
settings and “DONE” will be displayed. To
abort changes, press the Enter button to
re-enter the “Stanby Mode.” Other
attributes (KP, KI, KD) cannot be edited.
6
Stirrer Ready Cook-Down Temp. Hold Temp. Cook Dwell F° or C° Altitude
Max.
100% On/Off 207°F or (97°C) 140°F or 60°C 60 Minutes F°/C° +255' x 100
Min.
15% On/Off 190°F or (88°C) 125°F or 52°C
1 Minute F°/C°
– 14' x 100
(+ 25,500')
(
–
1,400')
^
^
Display
“STR” “RDY” “SCT” “SHT” “SCD” “DEG” “ALT”
*
* Default settings