46
Read/write registers
To set and operate the BOOSTIK, click on the
Control
tab in the left side half of the
Generic UI window to view the read/write registers. For read/write registers not
described, refer to the SDK instruction manual included with the SDK installation.
Figure 28
Read/write registers under the
Control
tab
Output power
setpoint
Input the optical output setpoint power level in mW or dBm. For setting the power
using mW units, use register 22, and for using dBm units, used register A0.
Figure 29
Setting the output power setpoint
Emission
Keying 1 into register 30 enables emission from the amplifier module. If there is no
input power (
), then enabling emission fails and a fault is
raised with error code 60 displayed – see
“Errors and status indicators” on page 50
Readings