7
Installation and Connection
7.1 Installation Instructions
-
Observe appropriate installation!
-
Follow applicable legal or operational guidelines!
-
Inappropriate use may result in injuries and/or damage on instruments!
The unit’s mounting place must be selected according to certain criteria.
Strictly avoid:
-
direct sunlight (use weatherproof cover if necessary, e.g. NIVUS Art. No.
ZMS01800)
-
objects radiating strong heat
-
permissible ambient temperature: -20° C to +40° C
-
objects with strong electromagnetic fields (frequency converters, high volt-
age power lines or similar)
-
corrosive chemicals or gases
-
mechanic shocks
-
vibrations
-
radioactive radiation
-
direct installation close to footpaths or travel ways
7.1.1 Fastening Wall Mount Enclosure
For fastening the wall mount enclosure, depending on place of mounting, use
4 machine screws size M5 with suitable length as well as the necessary nuts
and shims. Or use 4 wood screws, min. diameter 4.5 mm (0.1772 in). These
screws must penetrate min. 40 mm (1.575 in) into the wall or min. 50 mm
(1.97 in) into appropriate dowels.
The clear view door as well as the display of the measurement transmitter are
covered with a protection foil for protection during transport and from scratches
during assembly. This protection foil must be removed immediately after the as-
sembly!
Note
If the view door and display protection foil have been exposed to direct solar
radiation for a long period, they cannot be removed easily.
This problem may be solved by using methylated spirit or car polish if required to
clean the clear view door or the display. If these measures should not be suc-
cessful, clear view door as well as display can be replaced by NIVUS (subject to
costs).