Instruction manual
OCM Pro CF
®
OCM Pro CF - Rev. 09 as of 07.06.2016
page 141
h:
Distance from piezo crystal to water surface
2 >>> 1: Currently active piezo crystal
0,8 m:
Channel height
Fig. 9-136 Echo profile screen with sensor type DSM
The screen shown above indicates the current level in the channel at a mounting
height of 0.862 m – 0.776 m = 0.086 m.
T-Sensor
This screen allows to view the measured water and air temperature. The air
temperature measurement only possible by using one external air-ultrasonic
sensor type OCL/DSM driven by OCM Pro.
Invalid values indicate cable break, short circuits or incorrectly clamped
connections.
Fig. 9-137 Temperature screen
9.6.6 I/O Menu "Interfaces"
The menu includes the transmission speed of the internal interfaces. It has no
meaning for users and is going to be used for service purposes only.
9.6.7 I/O Menu "Controller"
This menu is visible only if the controller has been enabled in the PAR menu.
Having the controller enabled will bring up the screen below:
Fig. 9-138 Select controller information