Instruction manual
NivuSonic Clamp on
8.6.2 I/O Menu „Digital outputs”
This submenu indicates the calculated conditions to be output on the relay. We
distinguish between logically "OFF“ or "ON“.
Fig. 8-61
Screen digital values
The actual currents fed to the digital output clamps will not be indicated. The
only signal visible is the signal the analog output converter receives for output.
This menu cannot be used to detect or to indicate faulty wiring.
8.6.3 I/O Menu "Sensors"
This menu, including the respective submenus, allows to view and to asses the
most important sensor conditions. It hence provides information on the quality of
the measurement place, cable layout and many more parameters.
Fig. 8-62
Basic selection menu
T-Sensor
This screen allows to view the water temperature based on the measured ve-
locity of sound. Invalid values indicate faulty geometry settings (mounting posi-
tion) or incorrectly clamped connections.
Fig. 8-63
Temperature screen
LDV_Sensors
This submenu indicates the investigated flow velocities of the single paths.
The overall velocity required for flow calculation will be derived from these sin-
gle velocities.