On-time (select an impulse duration between 0.01 and 2.0 seconds. The ra-
tio between impulse and break is 1:1. It is useful to increase the impulse du-
ration beyond the 0.5 sec. default setting if slow PLC inputs or mechanic
counters are in use).
-
Amount (defines the impulse weighting. The measured amount is going to
be integrated internally until the selected value has been equalled. An im-
pulse signal with the previously programmed duration will be emitted after-
wards and the integrated internal value is going to be set to 0 again. This
procedure will restart subsequently.)
8.5.6 Parameter Menu "setup parameter"
Fig. 8-35
Setup parameter – submenu
This menu allows to either modify or to reset the following basic settings of the
system.
Load factory setup
Enables a general reset of the measurement transmitter. Selecting this point will
bring up the screen below:
Fig. 8-36
Executing a general reset
Selecting “YES“ is going to reset the system to the basic parameter settings.
The default settings are going to be restored and all customer modifications
will be set to default condition (system general reset.)
NivuSonic CO - Rev. 00 as of 08.12.2010