TECHNICAL DATA
3.1
G
ENERAL DATA
T
YPE
PDF-
01-2
PDF-401-A
E
X
PDF-401-C
E
X
PDF-
01-6
E
X
PDF-601-A
E
X
PDF-601-C
E
X
P
F-
01-4
P
F-401-B Ex
P
F-401-D
E
X
P
F-
01-8 Ex
P
F-601-B Ex
P
F-601-D
E
X
Powering 2-wire
3-wire
Input
Current loop
Range
3.6 … 22 mA
0 … 22 mA
Display/indication
6-digit LCD, engineering units and bargraph
Range of displayed value
-9999 … 29999
Engineering units (programmable)
mA, ºC, ºF, %, m, cm, in ,ft, l, m
3
, gal, ft
3
Displayed process values (programmable)
DIST
(distance),
LEV
(level),
VOL
(volume/content)
Ambient temperature
-40ºC…+70ºC, with display module: -25ºC…+70ºC
Accuracy
± 0,1 % for displayed values over 10000
Temperature coefficient
± 0.05 % / 10°K
Output ----
4…20 mA and/or HART
for 4…20mA current limit values: 3.9…20.5 mA
terminal resistor for HART R
tmin
= 250
Ω
Error indication on the current output
----
Output = 3.8 or 22 mA
Power supply
----
Standard version: 10 V…36 V, Ex version: 10 V ... 30 V
Load resistance
----
R
t
≤
(Us-10V) / 0,022A (Us = supply voltage)
Voltage drop
< 1.6 V
< 1 V
Overvoltage capability
140 mA
140 mA
Damping
3 s, 5 s, 10 s or 20 s
Electric protection
Class III
Ingress protection
IP67
Electric connection
Through M20x1.5 cable gland, shielded, twisted cable of
∅
6…12 mm, Ex version: see additional data for Ex certified versions
Housing
Paint coated aluminium PBT glass fibre reinforced or stainless steel (w. Nr 1.4571)
Mass
Aluminium housing appr. 0.9 kg, plastic housing appr. 0.55 kg, stainless steel appr. 2.4 kg