4 / 5
eka7021a0600h_07
Pipe extension Ø10x1
EAR–70
–1
Rigid pipe for cable extended version
EAK–7
–1
3- blade extended paddle (268 mm, 10.5 inches)
EAL–708–1
Mounting plate
EAM–70
–0
The extension can only be used with a
standard model.
It is crucial it the paddle of pipe-extended
models is not exposed to strong lateral forces,
for they may damage the pipe!
If the required “L” length is different than the
one in the order code, it must be specified.
By pulling it over the cable rope, it makes
units with cable rope extension stiffer.
Caution! If the cable rope is shortened, the
tube must be cut to a matching length.
The length of the exposed cable rope section
must be at least 30 mm (1.2 inches).
Use a mounting plate to insert paddle into the
tank!
Primarily for light materials due to its large blade.
Extended variant.
Paddle pressure*: 0,3 kPa (0.044 psi)
Stainless steel and carbon steel variant
available. The latter is electrostatic powder-
coated to prevent rusting.
Sliding sleeve for rod extended version
EAM–703–0
Flexible Coupling
EAS–701–0
Only for the E
R–7
variant, for changing
the insertion length.
To protect the paddle from strong dynamical
effects that may occur during operation.
3. INSTALLATION
The device can be installed in several ways, depending on the design (vertical, horizontal, angle), with threaded connections, mounting plates, adapters, adjusting units,
etc. The unit must be protected against steady material influx by selecting the appropriate mounting position or using an overhead protective shield. When the device is
mounted onto the side of the tank, coning or arching of the material must be taken into consideration. The preferred location for the unit is the one with the least amount of
vibration. It is important that the device is installed so that the cable glands are in the correct position.
If the device is mounted onto the side of the tank, deposits may clog the connection point and block the movement of the paddle; therefore, they must be prevented.
If a 3-blade unit is on a mounting plate, the insertion hole must be at least
110 mm (4.3 inches) for the paddle to pass safely. Do not forget to use the mounting plate
seal! When mounted on a bracket or mounting plate, a 1" or 1½" flat nut can be ordered to fasten the unit. Mounting plates thinner than 9 mm (0.35 inches) requires
spacers. Attaching rod extensions longer than 0.5 m (20 inches) to the tank wall is recommended. For upper-level switching, using a standard model with the appropriate
paddle between the process connection and the measured medium standard is recommended, and if necessary, it must be equipped with a flexible coupling or a rod
extension to reach the required detection level. When using the unit for low fail-safe indication, it is necessary to install a deflector element. A cable-extended version
recommended for low-level switching, with a suitably-sized rigid tube, if necessary. It must be installed vertically. It will not work correctly if it is slanted or horizontal. A
rigid tube is recommended if the measured medium forms a steep cone or the surface is so rigid that the paddle cannot penetrate the material, and the level tends to
deviate significantly from its vertical position. If so, it is recommended to use a rigid tube for proper operation.
Besides, there are countless ways to use the device, but you should always be careful about choosing the right process connection and protection of the device.
Protection with a protective shield
Connection for mounting plate (recommended)
Side mounting
Correct
Incorrect
Mounting
Correct
Incorrect
In case of side-mounted units, the cable glands should face
down!
*
The paddle pressure value is understood for 0.2 Nm (0.148 ft-lb) torque in the case of complete
immersion. The specified values are for comparing the paddles. Paddles with lower paddle pressure are
for lighter materials, while paddles with higher paddle pressure are recommended for heavier materials.