Quality Control Testing
To assure
you are getting accurate and
reliable results, TRUEresult twist 2 offers
two kinds of quality control tests. These
tests let you know that your TRUEresult
twist 2 System is working properly and
your testing technique is good.
Automatic Self-Test
An Automatic Self-Test ensures that
the Meter is working correctly.
1.
Wash hands and dry thoroughly.
2.
Remove one Test Strip from the
Test Strip vial and insert Test Strip
into the Meter.
3.
Meter turns on. The full Display
appears and is replaced by the
blinking Drop Symbol. The Meter
is working correctly and is ready to
perform a Control or blood test.
Do not use Meter if:
• The full Display does not appear
(segments are missing),
• The blinking Drop Symbol does not appear, or
• An error message appears in the Display.
See
Troubleshooting
or call for assistance (see cover
of this Booklet for phone number).
15
Drop Symbol
Full Display