
SAFETY PRECAUTIONS-3
ܭ
WARNING
z
Electronic equipment producing radio waves should not be used in the same room or building as
where this unit is used.
•
Avoid the use of electronic equipment that produces radio waves including cellular phones,
transceivers, and radio-controlled toys. Use of this equipment inside of a hospital building can
lead to malfunction of the machine.
•
To prevent a possible accident, inform and instruct people not to carry or use radio-wave
producing equipment in the hospital building.
z
Do not change the setting values during treatment if there is no emergency.
•
Change setting values, if any, before connection to the patient.
•
Before selecting a function, ensure to understand the function thoroughly.
z
Provide a drain opening in the floor.
•
The power cable should not touch the floor to prevent from electric shocks due to a technical
failure and liquid leakages.
•
Provide a drain opening near to the unit.
z
Make sure to read this Manual before starting maintenance and inspection.
•
Read and understand this Manual thoroughly before starting maintenance and inspection.
•
Exactly follow the described procedures for maintenance and inspection.
•
Maintenance and inspection out of a range described in this Manual must be entrusted to
specialists.
z
Do not choose unnecessarily functions during the treatment.
z
Do not incline the equipment by more than 5 degrees with something placed on the tank base or
hanger. Doing so may topple over the equipment.
z
Do not relocate the equipment with something placed on the tank base or hanger.
z
Do not tilt the unit more than 10 degrees. It can overturn.
z
Equipment, not permitted to be used together with this machine, may not be connected. The
total electric leakage current from the unit and other equipment together can exceed the
maximum allowable value.
z
The operation and maintenance not specified by the manufacturer and the application of
unspecified accessories may cause death or injury of the patient. The manufacturer is not
responsible for the patient safety when the unit is operated, maintained, or calibrated by
methods not specified by the manufacturer. Trained and authorized personnel must operate the
unit in accordance with the methods established by the manufacturer.
z
Check the dialysate parameters with external testing equipment before starting any treatment.
•
Verify the actual dialysate concentration by an osmometer, conductivity meter, flame photometer,
etc.
•
Check if the displayed concentration level is accurate.
z
Verify, at the end of the cleaning, that there is no residual chemical or acid present.
•
Verify, at the end of the disinfection the presence of chemicals (sodium hypochlorite, citric acid,
peracetic acid, or other chemicals) using a test paper or reagent.
Summary of Contents for SURDIAL
Page 2: ......
Page 4: ...PREFACE 2...
Page 16: ...CONTENTS 8...
Page 18: ...1 2...
Page 19: ...2 1 2 CONTRAINDICATION Any application other than dialysis using SURDIAL is contraindicated...
Page 20: ...2 2...
Page 22: ...3 2...
Page 24: ...4 2...
Page 25: ...5 FLOW CHART...
Page 26: ......
Page 32: ...5 6...
Page 33: ...6 DISASSEMBLING AND REPLACEMENT...
Page 34: ......
Page 40: ...6 1 1 6...
Page 44: ...6 1 2 4...
Page 48: ...6 1 3 4...
Page 50: ...6 1 4 2...
Page 54: ...6 1 6 2...
Page 56: ...6 2 1 2...
Page 60: ...6 2 2 4...
Page 64: ...6 2 3 4...
Page 66: ...6 2 4 2...
Page 72: ...6 2 5 6...
Page 73: ...6 2 6 1 6 2 6 Blood pump Assy BP BPC Fig 6 2 6 1 Blood pump Assy...
Page 76: ...6 2 6 4...
Page 84: ...6 2 7 8...
Page 85: ...6 2 8 1 6 2 8 Venous clamp Assy CLV Fig 6 2 8 1 Venous clamp Assy...
Page 87: ...6 2 9 1 6 2 9 Syringe pump Assy SP Fig 6 2 9 1 Syringe pump Assy...
Page 92: ...6 3 2 2...
Page 94: ...6 3 3 2...
Page 96: ...6 3 4 2...
Page 100: ...6 3 5 4...
Page 106: ...6 3 7 4...
Page 114: ...6 3 10 2...
Page 132: ...6 3 16 2...
Page 133: ...7 CONSTRUCTION OF ELECTRICAL PART...
Page 134: ......
Page 141: ...8 CONNECTION REPLACEMENT AND ADJUSTMENT...
Page 142: ......
Page 186: ...8 12 8...
Page 195: ...9 MAINTENANCE MODES...
Page 196: ......
Page 198: ...9 1 2...
Page 204: ...9 3 2...
Page 220: ...9 9 6...
Page 230: ...9 12 2...
Page 232: ...9 13 2...
Page 234: ......
Page 246: ...10 3 6...
Page 262: ...10 8 2...
Page 264: ...10 9 2...
Page 268: ...10 10 4...
Page 270: ...10 11 2...
Page 280: ...10 15 4...
Page 290: ...10 17 2...
Page 292: ...10 18 2...
Page 298: ...10 21 2...
Page 300: ...10 22 2...
Page 303: ...11 DIP SWITCHES...
Page 304: ......
Page 306: ...11 1 2...
Page 312: ...11 2 6...
Page 313: ...12 VARIOUS DATA...
Page 314: ......
Page 316: ...12 1 2...
Page 334: ...12 3 12...