21
2.
Press ▲ or ▼ to adjust the measuring mode, and
press
MAIN
button to confirm it.
General Tests (Gen) - any time of day without regard to
time since the last meal.
AC - no food intake for at least 8 hours.
PC - 2 hours after a meal.
QC - testing with control solution.
3.
Obtain a blood sample.
Use the preset lancing device to prick your desired site.
After penetration, discard the first drop of blood with a
clean tissue or cotton. Gently squeeze the punctured area
to obtain another drop of blood. Be careful NOT to smear
the blood sample. The volume of blood sample must be
at least 0.5 microliter (
μ
L) for blood glucose or at least
0.8 microliter (
μ
L) for β-Ketone.
Perform a Blood Glucose or β-Ketone Test
1.
Insert the test strip into its slot. Wait for the device to
display the test strip “ ”, blood drop “ ” and
GLU
or
KETONE
.
Summary of Contents for GlucoKey Connect TD-4183D/GD82a