11
Test Strip
1
2
Front of Test Strip
➀
Contact End - Insert into Meter with Contacts
(metallic blocks) facing up before adding sample.
➁
Sample Tip - Touch sample (blood or Glucose
Control) to edge of Tip.
Placement of Sample to Test Strip
Correct Incorrect
• Allow sample (blood or Glucose Control
drop) to be drawn into Sample Tip.
Caution!
Holding the Test Strip Sample Tip to the
blood sample too long after the Meter
begins testing may cause inaccurate results.
Note:
Do not apply blood or Glucose
Control to top of Test Strip.
• Do not smear or scrape drop with Test Strip.
• Do not apply more sample to Test Strip
after removing from drop.
Note:
Do not insert Sample Tip with
sample into Meter for testing.
May damage meter.
Contacts
Face Up