How to Test Control Solution
Use ONLY TRUE METRIX Control Solution with
the TRUE METRIX Self Monitoring Meter and Test
Strips.
1. Check dates on control solution
label and test strip vial label. Do not
use control solution or test strips
if expiration dates have passed.
(Control solution - 3 months after
first opening or date next to EXP on
label; test strips - 4 months after
first opening or date next to EXP on
labels.) Discard expired products
and use new products.
2. Allow control solution, vial of test strips and
meter to adjust to room temperature. Write date
first opened on both control solution bottle label
and test strip vial label when using for the first
time.
3. Gently swirl or invert control solution bottle to
mix.
DO NOT SHAKE!
4. Remove one test strip from vial. Close test strip
vial immediately. Use test strip quickly after
removal from vial.
17
May 30, 2015
TMX_OB_GutsENG_RE4NPD03r6.indd 17
7/15/15 3:37 PM
Summary of Contents for TRUE METRIX
Page 54: ...52 RE4NPD03 Rev 6 EN Notes ...