13
Operation
Transducer Tester
In the main menu screen, press
or
to select “2. TRANS TESTER”,
press
to enter “TRANS TEST SET”, press
, , or to change
the cursor location and then press or to
set the parameters of
start frequency, end frequency and step.
When testing the transducer, please set the scan range near the
resonance point. For example, when there is a 1000kHz transducer,
the range of scan settings should be 900kHz
~
1200kHz.
And then, press to start transducer test. During testing, press
to pause/redo the frequency scan. And press
to return back
the higher level menu.
After finishing the scan, the resonance is shown on the screen.
Note: After return back the main menu, the settings will be
returned to factory default.
“
Sfreq
”
is start frequency of test scanning.
“
Efreq
”
is end frequency of test scanning.
“Step
”
is step of frequency change every
time.
Factory default is shown in the left default.
“
Freq
”
is resonance frequency.
“
Zmag
”
is resonance impedance.
“Zphase
”
is
resonance phase
.
Factory default is shown in the left default.
“
Freq
”
is scanning frequency.
“
Zmag
”
is impedance of present frequency.
“Zphase
”
is
phase of present frequency
.