![Ningbo CM-1900C User Manual Download Page 23](http://html1.mh-extra.com/html/ningbo/cm-1900c/cm-1900c_user-manual_1694613023.webp)
23
equipment to avoid electromagnetic interference base on the equipment’s max rated power according to below
recommandation
Transmitter max
rated power(W)
W
isolation distance according to difference frenquency of the transmitter (m)
150kHz - 80MHz
d = 1.2
80MHz - 800MHz
d = 1.2
800MHz - 2.5GHz
d = 2.3
0.01
0.12
0.12
0.23
0.1
0.38
0.38
0.73
1
1.2
1.2
2.3
10
3.8
3.8
7.3
100
12
12
23
For the max rated power value is not listed in above form, user can use the distence (unit m), and use the correct
equation according to the emmision frequency to calculate; P is the max rated power, unit W.
Note: 1. For the frequency 80MHz and 800MHz, choose the equation for the higher frequencies.
2. this instruction may not apply for all situatyions. Propagation Coupling might be influenced by the absorbe
and reflection of buildings, objects and human bodies.
After sales service,If there is any problem with the device during usage and the
problem can not be solves after communication with the distributor, please fill in
below information and send to MSOC distributor:
1.
Product modle and name: CM-1900P, CM1900C, VX22
2. Product serial number(provided on the product label)
3. Description: detailed problem description of the problem