25
5
Lighting area adjustment (only for G-LS)
It is possible to adjust the lighting area by moving the lamp socket back
and forth. Adjust the lighting area to allow easy viewing.
6
Attaching filters (only for C-DSLS)
Rotate and remove the filter cap at the end of
the lamp house and insert a filter. Use filters
that have a diameter of 33 mm and a width of
1.5 ~ 2 mm.
Antistatic Feature
The SMZ645, the SMZ660, the C-PS/C-PSC plain focusing stand and the
C-W 10X eyepieces are made antistatic. You can enjoy the benefits of this
feature when observing the sample that cannot stand much static. When
using the microscope on this purpose, do not forget to change the stage
plate to antistatic ESD stage plate, and ground the microscope through the
ground jack at the rear of the focusing stand. The
φ
4 mm grounding
terminal can be inserted in the ground jack.
The following accessories are also made antistatic.
•
C-FPS Fluorescence Attachment
•
Auxiliary Objectives (all types)
•
C-FMB Focusing Mount B
•
C-US1 and C-US2 Universal Table Stands
Airtight Feature
The SMZ645, the SMZ660 and the C-W 10X eyepieces are made airtight.
This feature may be convenient when using the microscope in the oil-
misty environment in the factory or for soldering.
Note) Always attach the rubber cap on the rear of the zooming body and
the rubber cover on the eyepiece to hold the airtightness.
Using Accessories
IV
Filter
ø33 mm
Filter cap