7. Replacing the Lamp
Note the following before replacing the lamp in case that the lamp is out of
order.
•
Before replacing the lamp, turn off the power and unplug the power
cord.
•
After the power cord has been unplugged, wait for approximately 15
minutes or more for the lamp to cool down.
•
Use JCR 12V 20W/G1/A1.
•
Never use a LS-3510AF lamp (12V 35W). It may cause damage to the
unit.
•
Do not touch the bulb section of the lamp. Oil from your fingers will
shorten the life of the lamp.
The usual life of lamp used in the LS-4500AF is 2,000 hours. With normal
use, the lamp will need to be replaced approximately once every two years.
Step 1
Wait for about 15 minutes for the lamp to cool down after the power cord
has been unplugged, then open the cover on the top of the unit.