Chapter 8 Troubleshooting
4 Camera Settings on the PC
- 35 -
4
Camera Settings on the PC
Trouble Checkpoints
Unable to control
camera functions. (Or
incorrect settings
result.)
•
Check that the specified device driver is installed on the PC.
The specified device driver can be found on the supplied CD-ROM.
•
Check that the application software in use is compatible with the DS-U3.
•
Check that the DS-U3 and the PC are connected with the IEEE1394b
cable.
•
Check that the cable used is IEEE1394-compliant. For shielded twisted
pair cables, the cable connecting devices must be 4.5 m maximum.
•
Check that the version of the OS is appropriate for the product. The
product is not compatible with the Macintosh.
Applicable OS: Windows 7 Professional, Windows XP Professional
Service Pack 2 (SP2)
Noise is heard from the
Camera Head (DS-Ri1)
•
If you obtain a high-resolution image by pixel-shifting on DS-Ri1, noise
may be heard from the movable part of the Camera Head. However, this
is not a problem.
Adjusting
the
white
balance gain setting
seems to have no
effect on the image.
•
If you are using the GainUp mode (X 16) for DS-Ri1, be sure that the
white balance gain is set to below 4. Setting the parameter to 4 or higher
is invalid due to the hardware limitation.
Summary of Contents for DS-Fi1
Page 2: ......
Page 8: ...Safety Precautions vi ...
Page 16: ...Chapter 3 Names of Parts and Their Functions 8 ...
Page 26: ...Chapter 5 Microscope Setting 18 ...