To stop the calibration procedure prematurely, push the EC or pH button during calibration. 

 

The instrument can be set to factory calibration by pressing the "

" and "

" buttons together with the 

Temp-button during calibration. 

 

7  SWITCHING THE INSTRUMENT OFF 

The instrument can be switched off manually as follows: 

 

 
 
 
 
 
 
 
 
 

1. Keep the EC and pH-buttons  

2. Release the buttons when 

pressed. 

"OFF" is shown on the display. 

 

The instrument will switch off automatically after 90 or 180 seconds. The time can be adjusted as follows: 

 
 
 
 
 
 
 
 
 
 

1. Keep the "

" and "

" buttons 

2. The display shows the set time  3. Keep the "

" and "

" buttons pressed 

pressed. 

in seconds with the time  

until the desired time is shown. 

 

adjustment indication. 

8  ADJUSTMENT OF THE REFERENCE TEMPERATURE 

The temperature compensation of the EC measurement is always referring to a standard temperature. 
This value differs per country and per branch. Because of this the reference temperature is adjustable. 
Standard it will be 25°C (other options are 18°C or 20°C).  
Only the reference temperature of the measurement is adjustable. During calibration the reference 
temperature is fixed to 25°C. The reference temperature adjustment is as follows. 

 
 
 
 
 
 

 

1. Keep the temp and EC buttons  2. Release the buttons when 

3. Now "r 25" will appear on the 

4. Adjust with the 

 or 

 

at the same time pressed. 

“cal” is written on the display

 

display. This means that you are  button the right reference 

 

the two LED’s will keep alight

 

in the calibration procedure. 

Temperature. 

 
 
 
 

 

1. Confirm with the EC or the 
 Temp button. 

 

 

Summary of Contents for EP1400

Page 1: ...USER MANUAL EP1400 0 TEMPERATURE COMPENSATED EC pH C METER WITH AUTOMATIC CALIBRATION Valid from S N 075663...

Page 2: ...DESCRIPTION 4 2 TECHNICAL SPECIFICATIONS 5 3 MEASURING WITH TEMPERATURE COMPENSATION IN A LIQUID 6 4 MEASURING WITHOUT TEMPERATURE COMPENSATION IN A LIQUID 7 5 MEASURING EC AND pH IN SOIL 7 6 CALIBRA...

Page 3: ...EP1400 it is very easy to measure EC pH and temperature in liquids and soils The instrument is supplied in a useful carrying case complete with calibration liquids and all the accessories needed for...

Page 4: ...O 2 3 4 6 5 7 8 9 11 12 13 BATT 1 Display 11 Time adjustment indication 2 Temp adjustment time adjustment 12 Stability check indication 3 Temp adjustment time adjustment 13 Low battery indication 4 EC...

Page 5: ...7 1 00 pH Recognition pH 4 90 105 2 Points calibration with automatic recognition of standard buffer solutions pH 4 pH 6 86 and pH 9 Possibility for manual calibration with other buffer solutions 3 T...

Page 6: ...ement 1 Push the pH button The LED on 2 Remove the bottle rinse 3 Put the EC and pH electrodes in 4 Wait until the value is stable the button will blink the electrode in demiwater the liquid and rinse...

Page 7: ...pter 3 5 MEASURING EC AND pH IN SOIL For measurements of EC and pH in soil firstly we have to make a soil sample For an extraction of 1 2 the next steps has to be followed Fill the ABC sample cup up t...

Page 8: ...stable it 8 When the value is recognised buffer solution and rinse carefully sign is visible the meter will be compared to the standard it appears on the display checks the stability values 9 When thi...

Page 9: ...standard values liquid with the same temperature as the pH 8 86 buffer solution 9 When the solution is recognised 10 When this is the right value 11 PH2 will appear on the display 12 Clean the pH elec...

Page 10: ...perature or EC button during calibration The instrument can be set to factory calibration by pressing the and buttons together with the pH button during calibration Temperature calibration 1 Keep the...

Page 11: ...sed pressed in seconds with the time until the desired time is shown adjustment indication 8 ADJUSTMENT OF THE REFERENCE TEMPERATURE The temperature compensation of the EC measurement is always referr...

Page 12: ...r messages can appear on the display ERR3 The calibration of the electrode is out of range 70 100 Press the EC button to stop the error message During 10 seconds one of the following messages will app...

Page 13: ...des after measurement always with distilled water Protect the pH electrode against desiccation by storing it always in storage liquid Protect instrument and electrodes against high shocks and dirt 11...

Page 14: ...EP1400 EC pH C METER _____________________________________________________________________________________________ 14...

Page 15: ...EP1400 EC pH C METER _____________________________________________________________________________________________ 15...

Page 16: ......

Reviews: