ALARMS / SAFETY DEVICES 

 

Power Failure Alarm

 – In the event of a loss of power, a continuous audible alarm is activated. 

 

Process Failure Alarm

 – In the event of a process failure, the Red Light Indicator (6) and audible alarm is activated. 

 

Electrical Protection Device

 – in the event of a surge or drop in the power supply the Circuit Breaker (5) will trip.   

Restart the unit by depressing the reset button on the Circuit Breaker (5)   

 

Green Light Indicator

 (6) indicates that the unit has power and is ready to provide oxygen enriched air to the patient. 

 

Red Light Indicator

 (6) indicates that a system fault has occurred or there is no power to the unit. 

 

SPECIFICATIONS 

Flow Values: 

2 to 8 Liters per Minute 

Characteristics: 

Electrical: 

 

Length – 15.5” or 394 mm 

120 VAC, 60 Hz, 4.0 amps, 490 watts

 

Average Oxygen Content: 

Width – 15.6” or 396 mm 

230 VAC, 50-60Hz, 2.0 amps 490-585 watts 

 at 2 Liters per Minute – 93% 

Height – 27.8” or 706 mm 

Two-Prong polarized plug 

 at 6 Liters per Minute – 92% 

Weight – 54 lbs. or 24 kg. 

Double Insulated Cabinet 

 at 8 Liters per Minute – 90% 

(Varies with models) 

 

 
Storage Instructions: 

This device should only be stored, transported and operated in the vertical position. 

 

Disposal of Device and Accessories: 

The unit and accessories (tubing, humidifier, etc.) should be disposed of using the methods 

appropriate to the local waste authority where used. 

 

TROUBLESHOOTING

 

OBSERVATIONS 

POSSIBLE CAUSES 

SOLUTIONS 

The (I/0) Power Switch is in the (I) On 
Position but the device does not operate.  
The audible alarm sounds continuously 

Power Cord (10) is not 
plugged into outlet 
 
Power failure 

Check that the unit is properly plugged into the electrical outlet 
 
Check the Circuit Breaker (5) and reset if necessary 

 
Red Light Indicator (6) remains lit 

Product pressure or purity is 
not at an acceptable level 

Contact your Equipment Provider 

 
The audible alarm does not sound when 
unit is first turned on 

Faulty 9 volt battery 
 
Internal electrical fault 

Contact your Equipment Provider 

 
The unit is operating but the Green Light 
Indicator (6)  is not lit 

Faulty indicator 

Contact your Equipment Provider 

 
The unit is operating but there is no flow 
(flowmeter ball not moving) 

Internal system failure 

Stop device immediately and contact your Equipment Provider 

 
The unit is operating but the audible alarm 
sounds continuously 

Internal fault 

Stop device immediately and contact your Equipment Provider 

 
The unit suddenly stops and then starts 
again in a few moments. 

Dirty Filters 
 
Compressor Thermal Shut-Off 
 

Clean External Cabinet Filter (8) 
 
Contact your Equipment Provider 

The oxygen flow is interrupted or the flow 
at the cannula feels irregular 

Tubing is disconnected or not 
tight 
 
Tubing is restricted  

Check tubing connections 
 
Straighten tubing 
 
Contact your Equipment Provider 

 

 

NIDEK MEDICAL PRODUCTS, INC. 

3949 VALLEY EAST INDUSTRIAL DRIVE 

BIRMINGHAM, ALABAMA  35217   USA 

 
 