Nidek Medical Mark 5 Nuvo Lite Series Instructions For Use Manual Download Page 6

______________________________________________________________________________________________________________ 

2010-8401CE-E 

February 2018 

Page 6 of 8 

DO NOT disassemble due to danger of electrical shock. 
Refer servicing to qualified service personnel. 

The expected service life of this device is 10 years  with 
routine preventive and required maintenance. 

Preventive  Maintenance

Wash  cabinet  filter  (see 

“Cleaning  and  replacing  filters”)  weekly  or  after 
approximately 100 hours of use and for each new patient. 
More  frequent  cleaning  is  recommended  in  dusty 
environments.    Inspect  inlet  air  filter  (Fig  5)  at  each 
patient  visit.  Replace  filter  annually,  or  more  often 
depending on environment.  Check oxygen concentration 
every  15,000  hours  or  3  years  to  verify  the  continuing 
OCSI function.
The  manufacturer’s  instructions  for  the  preventive 
maintenance  of  the  devices  are  defined  in  the  service 
manual,  (Ref.  2010-8405).  Check  with  your  service 
provider for any updates to recommended schedules. The 
work must be carried out by suitably trained technicians 
certified  by  the  manufacturer.  Use  original  spare  parts 
only (see  “Accessories and Spare Parts”). Upon request, 
the  supplier  can  provide  circuit  diagrams,  spare  parts 
lists, technical details or any other information of use to 
qualified  technical  personnel  for  parts  of  the  device 

which  are  designated  as  being  the  manufacturer’s 
responsibility or by the manufacturer as repairable. 

DISPOSAL 

6.1 

Method for Waste Disposal 

All  waste  from  the  device  (Patient  Circuit,  Filters,  Etc.) 
must  be  disposed  of  using  methods  appropriate  to  the 
civil authority of the location where disposed. 

6.2 

Disposing of the Device 

This  device  has  been  supplied  by  an  environmentally 
aware manufacturer. A majority of the parts in the device 
are recyclable. 
Follow  local  governing  ordinances  and  recycling  plans 
regarding disposal of the device or components normally 
used  in  operation.  Any  accessories  not  original  to  the 
device  must  be  disposed  of  in  accordance  with  the 
individual  product  markings  for  disposal.  Furthermore, 
as  part  of  the  marking  directive  93/42/EEC,  the  serial 
number of the device disposed of must be sent to Nidek 

Medical if the unit has the   

    marking. 

TROUBLESHOOTING 

Observations 

Possible Causes 

Solutions 

The I-O (ON/OFF) button is in the “I” (ON) position but 
the device does not operate. 

Mains cable (Fig 2-8) is not correctly 
plugged into the wall outlet. 

Check the cable connection. 

The no voltage detection alarm test does not work. 
(See Alarms and Safety Features) 

Capacitor is not charged Internal 
electrical fault. 

Check the circuit breaker (Fig 1-4) on the front 
of the unit; Reset if necessary. 
Plug unit in for 10 minutes and retest. 
Contact your equipment supplier. 

The Power Switch (Fig 1-3) is in the “I” (ON) position, 
the compressor is operating and there is a flow but 
the green light is not lighted. 

Faulty indicator. 

Contact your equipment supplier. 

The Power Switch (Fig 1-3) is in the “I” (ON) position 
but there is no flow. 
The audible alarm sounds continuously. 

Pneumatic connection broken or 
other pressure problem. 

Stop the device by pressing the Power Switch 
(Fig 1-3) button. 
Contact your equipment supplier. 

The Power Switch (Fig 1-3) is in the “I” (ON) position, 
the compressor is operating and there is a flow but 
the audible alarm sounds continuously. 

Internal electrical fault. 

Stop the device by pressing the Power Switch 
(Fig 1-3) button. 
Contact your equipment supplier. 

Pneumatic circuit fault or low purity.

The compressor stops in mid-cycle, then starts again 
after a few minutes. 

Compressor thermal safety device 
has been activated. 

Stop the device and wait for it to cool down. 

Dirty Filters. 

Clean cabinet filter. Restart. If the device does 
not start, contact your equipment supplier. 

Cooling fan(s) not operating. 

Stop the device by pressing the Power Switch 
(Fig 1-3) button. 
Contact your equipment supplier. 

The oxygen enriched air flow is interrupted at the 
nasal cannula outlet. 

Tube disconnected or humidifier cap 
is not tight. 

Check that tubing connections are secure and 
that the humidifier is sealed. 

The flow at the nasal cannula outlet is irregular. 

Cannula tubing is kinked or 

restricted. 

Straighten the tubing. 
Contact your equipment supplier if damaged. 

Summary of Contents for Mark 5 Nuvo Lite Series

Page 1: ... flammable products on the device the administration accessory cannula or the patient s face neck Only persons who have read and understood this entire manual should be allowed to operate the device CONTRAINDICATIONS Those who continue to smoke because of the increased fire risk and the probability that the poorer prognosis by smoking will offset the treatment benefit Only Federal Law US restricts...

Page 2: ...e used when your settings were determined While undergoing oxygen therapy if you feel discomfort or experience a medical emergency seek medical assistance immediately The Mark 5 Nuvo Lite Family begins their operation with air being pulled into the external air intake filter This filtered air enters the compressor via a suction resonator and fine filter Pressurized air then exits the compressor an...

Page 3: ...pplied the device has a Blockage Alarm A continuous audible alarm and both indicator lights will be lit immediately in the event the flow of oxygen to patient becomes blocked Malfunction detection If low pressure occurs due to a mechanical failure the indicator light will flash yellow and a continuous audible alarm will actuate If any of the above alarm conditions occur press the Power Switch Fig ...

Page 4: ...mance and void the manufacturer s responsibility Spare Parts Part Ref Cabinet air filter Fig 2 9 8400 1025 Inlet Filter Under Fig 2 9 8400 1180 Ventilation Grill Fig 2 9 8400 0108 Casters 8300 8068 Mains Cable Wrap 8400 0022 Please consult the Nuvo Lite Maintenance Manual PN 2010 8405 for instructions on replacing any above spare parts 3 UNPACKING AND INSPECTION The Oxygen Concentrator is packaged...

Page 5: ...s or other by placing the orifice s on the surface of a glass of water The flow should disturb the surface of the water 6 Adjust the nasal cannula to suit your face 4 3 Shut Down At the end of the treatment press the Power Switch Fig 1 3 to the O OFF position to stop the device The oxygen enriched air flow continues for approximately one minute after the device is stopped Make sure during operatio...

Page 6: ...vidual product markings for disposal Furthermore as part of the marking directive 93 42 EEC the serial number of the device disposed of must be sent to Nidek Medical if the unit has the marking 7 TROUBLESHOOTING Observations Possible Causes Solutions The I O ON OFF button is in the I ON position but the device does not operate Mains cable Fig 2 8 is not correctly plugged into the wall outlet Check...

Page 7: ...______________________________________________________________________________________________________________ 2010 8401CE E February 2018 Page 7 of 8 EN 8 EMC INFORMATION ...

Page 8: ...eriodic servicing by an approved technician do not comply with the manufacturer s specifications the manufacturer is not responsible in the event of an accident or non performance This device complies with the requirements of the FDA Quality System Regulation and 93 42 EEC European directive but its operation may be affected by other devices being used nearby such as diathermy and high frequency e...

Reviews: