background image

The FG-7000L Series digital force gauges provide additional ten-

sion and compression testing flexibility with their external load cell 

input.  External load cells are immediately recognized when con-

nected to the display base.  Two styles of load cells are available 

for your specific testing needs.  The first is the ‘S’ Beam style load 

cell for both tension and compression tests with capacities rang-

ing from 220 lbf (100 kgf) up to 4500 lbf (2000 kgf).  The second 

style available is the ‘Ring’ type load cell for compression applica-

tions with capacities ranging from 2250 lbf (1000 kgf) all the way 

up to a hefty 44,000 lbf (200 kN).  

The multiple-language FG-7000L’s provide menu programming 

for easy selection and set-up of the instrument to your desired 

requirements.  Four selectable modes of operation include: Track 

mode for live readings, Peak mode for displaying the maximum 

reading, Auto Peak where the peak resets after a programmed pe-

riod of time and First Peak where only the initial peak is recorded 

once a decrease is sensed. The display has two selectable oper-

ations, numerical view with bar graph or graphical view with bar 

graph that if alarm tolerances are set, provides the user a quick 

view where their process is in relation to their upper and lower 

limit graph lines as well as pass/fail status. 

These high-tech instruments can easily data log a reading at the 

push of a button for simple data acquisition, or be set to continu-

ous data storage. Data can be viewed on the screen, sent to the 

optional printer, or loaded to be analyzed and graphed on the free 

software program. The 1,000 point memory with definable groups 

allows for multiple tests to be recorded and easily separated upon 

loading.  In addition, the comparator output can be set up for inte-

gration of the instrument into a quality system for repetitive testing 

such as on a production line.

The FG-7000L’s robust housings are designed to fit perfectly in 

the operator’s hand for portable testing. The large back-lit, 180° 

auto-reversible display, compression/tension icons, combined 

with the dual labeled key pad allows for usage of  the gauge in 

various positions while still being able to easily view and operate.  

These many features make the FG-7000L the ideal force instru-

ment for various applications such as, incoming quality inspec-

tion, finished goods testing, R&D or almost any force testing re-

quirement.

FG-7000L Digital Force Gauge
Operation Manual

SHIMPO INSTRUMENTS

Operators should wear protection such as a mask and gloves in case 

pieces or components break away from the unit under test.

Whether the unit is ON or OFF, DO NOT exceed the capacity of the 

sensor.  NEVER exceed 150% of the rated capacity for S beam, or 120% 

for ring type, or the load cell will be damaged.  At 110% of the rated 

capacity the display will flash a warning.

When mounting FG-7000L Series Digital Force Gauges, use M6 mount-

ing screws with a maximum insertion depth of 7 mm into the gauge. 

Hand tightens mounting screws, DO NOT use tools. Do not use dam-

aged clamp.

Measure in line tension and compression forces only. DO NOT attempt 

to measure forces at an angle to the sensor – damage to load cell may 

result.

Do not attempt to repair or alter this instrument.  Warranty will be voided 

and damage to the unit may result.

Use and store within the stated temperature and humidity ranges, or 

damage and failure may result.

If not using this instrument for extended periods of time, remove the 

batteries to prevent potential battery leakage from causing product 

damage.

DO NOT use tools to over torque the connection adapter. Hand-tighten 

only so damage does not occur.

SPECIFICATIONS

Accuracy:

 ± 0.2% F.S.

Selectable Units:

 kN, N, kgf, tf, lbf and klbf. (Depending on 

Range)

Overload Capacity:

 S beam: 150% of F.S.; Ring Type 120% of 

F.S. (LCD flashes beyond 110% of F.S.) 

Measurement method:

 Peak, Autopeak, First Peak or Track 

Mode 

Data Sampling Rate:

 1000 Hz

Display:

 160*128 dot matrix LCD with LED Backlight

Display Update Rate:

 10 times/second

Resolution:

 (See page 2)

Memory:

 1000 data

Set Point:

 Programmable high and low limits

Battery Indicator:

 Display flashes battery icon when battery is 

low

Power:

 3.6VDC 800mAH Ni-MH rechargeable batteries 

Battery Life:

 Approximately 16 hours continuous use per full 

charge

Charger / Adapter:

 Universal USB/BM charger, Input: 110 ~ 

240VAC

Temperature Effects:

 <0.054% per °F (0.03% FS per °C)

Outputs:

 USB,Serial Port RS-232, High & Low Limit NPN

Operating Temperature:

 14 to 104°F (-10 to 40°C)

Storage Relative Humidity: 

20 to 80%

Housing: 

Aluminum

Storage Temperature:

 -4 to 122°F (-20 to 50°C)

Oper. Relative Humidity:

 5 to 95%

Dimensions:

 5.7 x 2.9 x 1.4” (145 x 73 x 35.5 mm)

Product Weight:

 1.5 lb (0.7 kg)

Package Weight:

 2.8 lb (1.3 kg)

Warranty:

 1 year

Included Accessories:

 AC Adapter/Charger, USB cable, calibra-

tion cert.

LOAD CELL SPECIFICATIONS

Zero Balance: 

S Beam & Ring Type ±2% F.S.

Non-Linearity: 

S Beam 0.03% F.S.; Ring Type 0.1% F.S.

Hysteresis: 

S Beam 0.03% F.S.; Ring Type 0.1% F.S.

Temp. Effect: 

S Beam 0.03%/10°C F.S.; Ring Type 0.05%/10°C 

F.S.

Overload Protection: 

S Beam 150%; Ring Type 120%

Protection Class:

 IP76

www.

GlobalTestSupply

.com

Find Quality Products Online at:

[email protected]

Summary of Contents for Shimpo FG-7000L-S-5

Page 1: ...rs should wear protection such as a mask and gloves in case pieces or components break away from the unit under test Whether the unit is ON or OFF DO NOT exceed the capacity of the sensor NEVER exceed...

Page 2: ...0 5 0 01 S 20 20kN Capacity 2000 4500 20 Resolution 0 5 1 0 005 Ring Type for Compression N kgf lbf kN klbf tf R 10 10kN Capacity 10000 1000 2250 10 Resolution 0 1 0 1 0 5 0 01 R 20 20kN Capacity 2000...

Page 3: ...to cancel or touch MENU to confirm and exit Figure 4 2c Figure 4 2c 4 2 2 Group When several test samples need to be measured the sam ples can be coded into groups The range is 01 99 When set to 00 b...

Page 4: ...ENU to confirm and exit Figure 4 3b Single At the home screen pressing the LOG stores the cur rent displayed value If the default settings key is for storage See 4 5 8 key setting Series Continuous da...

Page 5: ...rs may be set per Fig ure 4 5a 4 5b Figure 4 5a Figure 4 5b 4 5 1 Display Mode Two display modes may be selected Digital and Graphic Fig ure 4 5c Figure 4 5c 4 5 2 Display Direction Select the mode of...

Page 6: ...can restore it to the factory default settings Carefully use this function Figure 4 5l Figure 4 5l 4 6 Language Select between English German and Chinese Figure 4 6a Figure 4 6a 4 7 Calibration Users...

Page 7: ...1a The USB Port can be connected to a charger with the USB cable for charging the internal Ni MH battery and can be connected to a computer for uploading the measured values Connect the gauge and the...

Page 8: ...ee Figure 6 1a 7 3 Dimensions S Beam Ring Type Capacity Dimensions W H B M 1kN 51 76 2 19 M12X1 75 2kN 5kN 10kN 25 4 20kN 76 2 108 M18X1 5 Capacity Dimensions mm D d L H SR DI M 10kN 32 8 14 16 16 24...

Reviews: