To turn the tracking function off, follow these steps:
1. Turn POWER off

2. Press PEAK and UNIT switches simultaneously and hold. Then 
press and release POWER. If the tracking function was on, the 
display will show TRK OFF momentarily to confirm successful dis-
able. 

3. Release PEAK and UNIT

The tracking function is now off. Repeat these steps to activate 
the tracking function.  The gauge displays TRK SET when reac-
tivated. 

FUNCTION SETTING
To change the default parameters of the gauge, follow the proce-
dure to access the settings:

1. Press and hold the ZERO button.

2. Press and release the POWER button

3. Each time the PEAK button is pressed, the unit will scroll through 
each of the functions.

4. Press the ZERO button to exit.

4

F01

0001

F01

-0001

F01

0001

UNIT

UNIT

UNIT

UNIT

UNIT

UNIT

UNIT

UNIT

UNIT

UNIT

MINUS

PLUS

PEAK

Power OFF; 

PEAK

F02

3

F02

5

F02

10

F02

20

F02

1

F02

2

F02

3

F03

10

F03

OFF

F03

10

UNIT

UNIT

UNIT

Display Update 
Time

Plus or Minus
Sign

Auto Power
Off

Press and hold ZERO button. Press and release the POWER button.

ZERO; 
Save and Exit

Compression Measurement
To measure compression forces, use the flat head attachment 
provided. The display will not show any sign for compression. 

Note: To display a minus sign (-) for compression forces, consult 
the “Function Settings” section for making the change.

Low Battery
When the battery charge is low, LO BAT will appear indicating 
the batteries need to be charged.  To completely charge fully de-
pleted batteries requires approximately 17 hours of charging time 
with the unit off. The units possess a smart charging system that 
shuts off the charging circuit once the batteries are fully charged. 

Note: The unit may operate with the charging adapter while the 
batteries are being charged. This will however lengthen charging 
time. 

Auto Power Off
If the gauge is on and there is no activity for 10 minutes, the unit 
automatically powers off to conserve the batteries. At 9 minutes 
without activity, PWR appears to notify that there is 1 minute be-
fore auto power-off. If the charging adapter is powering the gauge, 
the auto off function is inactive. 

Note: To modify the auto power off function, consult the “Function 
Setting” section.

Tracking Function
The tracking function accounts and compensates for temperature 
drift. It is advised to keep this function running for optimal per-
formance. However, when measuring very minute forces (a few 
ounces or grams) at a very slow rate, it may be advisable to dis-
able this function.

Summary of Contents for FGE-HXY

Page 1: ... the unit is detected do not operate the unit Notify the shipping carrier immedi ately to obtain damage claim instructions Only measure forces that are in line with the measur ing shaft Do not attempt to take any measurements tension or compression at any angle Failure to keep measurements in line will damage the instrument SPECIFICATIONS Capacity 500 lb 250 kg 2500 N Resolution 0 1 lb 0 1 kg 1 N ...

Page 2: ...E k PRODUCT DIAGRAM 2 ...

Page 3: ...l appear and be reversed 5 Release UNIT the display stays now in the reverse mode To return to the normal mode repeat steps 1 5 3 Select Average or Peak The unit can display the live Averaged reading or the Peak read ing Standard is Average If you want to measure the Peak force follow this procedure 1 Press PEAK the word PEAK will appear in the upper left corner of the display If you wish to have ...

Page 4: ...ment provided The display will not show any sign for compression Note To display a minus sign for compression forces consult the Function Settings section for making the change Low Battery When the battery charge is low LO BAT will appear indicating the batteries need to be charged To completely charge fully de pleted batteries requires approximately 17 hours of charging time with the unit off The...

Page 5: ...range calibration weight of your unit 3 Turn POWER OFF 4 Attach the fixture on the shaft of the force gauge Thread and tighten onto shaft by hand only 5 Press and hold the UNIT PEAK and ZERO switches 6 Press and release the POWER switch while still holding the other 3 switches Continue holding the 3 switches until the display reads CAL now release the three switches 7 Press the UNIT switch The dis...

Page 6: ...AWN CHECKED UNITS SCALE TITLE Toshi MO YS mm 1 1 MODEL ORIGINAL DRAWING NO DRAWING NO BGBM068 FGC H FGV H FGN H FGE H High Load Force Gauge Attachment Drawing M10 EJ350280 DATE 04 24 2017 Itasca IL 60143 USA Tel 630 924 7138 ...

Reviews: