6.1.2. Standard gas measurement is not reproducible.
Is standard sample
measurement
reproducible?
Is the syringe
defective? (Leaks,
needle clogging, or
contamination)
Is the septum
contaminated?
Is any carrier gas
leaking?
Locate the leak and
repair it.
Replace the syringe.
START
END
YES
NO
YES
NO
YES
YES
Replace the septum.
Is standard sample
measurement
reproducible?
YES
Wash the gas washing
bottle, dehumidifying bottle,
joint, and coupling.
Carry out maintenance of the PE-1.
Wash or replace the V5, joint, Teflon
tubes, or H5 collection tube.
NO
NO
NO