![NI SA-413T Manual Download Page 7](http://html1.mh-extra.com/html/ni/sa-413t/sa-413t_manual_1685461007.webp)
Item
Description
1 Connector
Connects to output connector of NI 4138/4139
Cleaning Statement
Caution
Clean the hardware with a soft, nonmetallic brush. Make sure
that the hardware is completely dry and free from contaminants before
returning it to service.
Environment
Maximum altitude
2,000 m (800 mbar) (at 25 °C ambient temperature)
Pollution Degree
2
Indoor use only.
Operating Environment
Ambient temperature
range
0 °C to 55 °C (Tested in accordance with IEC 60068-2-1 and
IEC 60068-2-2. Meets MIL-PRF-28800F Class 3 low temperature limit and
MIL-PRF-28800F Class 2 high temperature limit.)
Relative humidity range 10% to 90%, noncondensing (Tested in accordance with
IEC 60068-2-56.)
Storage Environment
Ambient temperature
range
-40 °C to 71 °C (Tested in accordance with IEC 60068-2-1 and
IEC 60068-2-2. Meets MIL-PRF-28800F Class 3 limits.)
© National Instruments
7
Triaxial Adapter for NI System SMUs Specifications