NI PXIe-4304/4305 Calibration Procedure

Connecting the TB-4304

The TB-4304 provides connections for the PXIe-4304/4305. Figure 1 shows the pin assignments 
of the TB-4304.

Figure 1.  

TB-4304 Circuit Board Parts Locator Diagram

Each channel consists of two terminal connections specific to that channel as shown in Table 2. 
You can verify or adjust the accuracy for any or all channels depending on the desired test 
coverage. Refer to Figure 2 and only connect the input channels required for verification or 
adjustment in parallel.

Refer to Table 2 for the analog signal names of the TB-4304.

Complete the following steps to connect the TB-4304.
1.

Install the PXIe-4304/4305 and the TB-4304 in the PXI Express chassis according to the 
instructions in the 

NI PXIe-4304/4305 and TB-4304 User Guide and Terminal Block 

Specifications

.

2.

Configure the PXIe-4139 to voltage output mode and enable remote sensing. Connect the 
PXIe-4139 output to the TB-4304 as shown in Figure 2.

Table 2.  

TB-4304 Analog Signal Names

Signal Name

Signal Description

AI+

Positive input voltage terminal

AI-

Negative input voltage terminal

AIGND

Analog ground input

NI TB-4304                                          

AI3-

AI3+

AI2+

AI4+

AI4-

AI5-

AI7-

AI7+

AI6-

AI6+

AI5+

AI0+

AI0-

AI1-

AI2+

AI1+

AI9+

AI9-

AI10+

AI13+

AI13-

AI14+

AI15+

AI15-

AI14-

AI12+

AI11+

AI10-

AI8+

AI11-

AI8-

AI12-

AI7+

AI7-

AI18+

AI21+

AI21-

AI22+

AI23+

AI23-

AI22-

AI20+

AI19+

AI18-

AI6+

AI19-

AI6-

AI20-

AI5+

AI5-

AI16+

AI29+

AI29-

AI30+

AI31+

AI31-

AI30-

AI28+

AI17+

AI16-

AI4+

AI17-

AI4-

AI28-

FOR PATENTS: NI.COM/PATENTS

S/N

158016B-02L

COPYRIGHT 2015

UL94V

-0

AIGND

AIGND

AIGND

AIGND

J2

1J

4

1

88

J5

J3

J6

J8

J9

J7

J10

J12

18

18

81

81

AIGND

AIGND

AIGND

AIGND

J15

J17

J11

J13

J18

J16

J14

8

1

8

1

U2

R2

U3

C1

U1

C3

C5

C2

R1

R5

R6

R4

R7

R8

R3

CR1

U4

C6

CR2

C4

PFI0

DGND

Summary of Contents for PXIe-4304

Page 1: ... 10 Specifications 10 World Wide Support and Services 10 Software Calibrating the NI PXIe 4304 4305 requires the installation of NI DAQmx on the calibration system Driver support for calibrating the PXIe 4304 4305 was first available in NI DAQmx 15 1 1 For the list of devices supported by a specific release refer to the NI DAQmx Readme available on the version specific download page or installatio...

Page 2: ... reference information NI PXIe 4304 4305 Specifications PXIe 4304 4305 specifications and calibration interval NI DAQmx Readme Operating system and application software support in NI DAQmx NI DAQmx Help Information about creating applications that use the NI DAQmx driver LabVIEW Help LabVIEW programming concepts and reference information about NI DAQmx VIs and functions NI DAQmx C Reference Help R...

Page 3: ...able 1 Table 1 Recommended Equipment for PXIe 4304 4305 Verification and Adjustment Equipment Recommended Model Requirements DMM PXI 4071 Use a DMM that has an accuracy of 32 ppm or better when measuring voltages higher than 10 V and an accuracy of 2 1 μV or better when measuring 0 V PXI Express Chassis PXIe 1062Q If this chassis is unavailable use another PXI Express chassis such as the PXIe 1082...

Page 4: ...I PXIe 4304 4305 and TB 4304 User Guide and Terminal Block Specifications 2 Configure the PXIe 4139 to voltage output mode and enable remote sensing Connect the PXIe 4139 output to the TB 4304 as shown in Figure 2 Table 2 TB 4304 Analog Signal Names Signal Name Signal Description AI Positive input voltage terminal AI Negative input voltage terminal AIGND Analog ground input NI TB 4304 AI3 AI3 AI2 ...

Page 5: ...d pair wire to eliminate noise and thermal offsets Maintain an ambient temperature of 23 C 5 C The PXIe 4304 4305 temperature will be greater than the ambient temperature Maintain an internal device temperature range of Tcal 5 C for the PXI 4071 1 Keep relative humidity below 80 Allow a warm up time of at least 15 minutes to ensure that the PXIe 4304 4305 measurement circuitry is at a stable opera...

Page 6: ...the sequence of operations and provides the test points required to verify the PXIe 4304 4305 The verification procedures assume that adequate traceable uncertainties are available for the calibration references The PXIe 4304 4305 has 32 independent analog input channels You can verify the accuracy of any or all of the channels depending upon your desired test coverage Complete the following steps...

Page 7: ...ement Range 1 V range for test points of 0 V 100 V range for all other test points Digital Resolution 7 5 digits Aperture Time 100 ms Autozero On ADC Calibration On Input Impedance 10 MΩ DC Noise Rejection High Order Number of Averages 1 Power Line Frequency Dependent upon the local power line characteristics Table 5 AI Voltage Mode Setup Configuration Value Channel Name Dev1 aix where x refers to...

Page 8: ...B 4304 as shown in Figure 2 3 Call the DAQmx Initialize External Calibration function with the following parameters Device In Dev1 Password1 NI 4 Call the 4304 4305 instance of the DAQmx Setup SC Express Calibration function with the following parameters calhandle in calhandle output from DAQmx Initialize External Calibration rangeMax Appropriate Range Max starting with the first Test Point value ...

Page 9: ... for the remaining Test Point values from Table 7 for the corresponding range that was configured in step 4 10 Repeat steps 4 through 9 for the remaining ranges from Table 7 11 Call the 4304 4305 instance of the DAQmx Adjust SC Express Calibration function with the following parameters calhandle in calhandle output from DAQmx Initialize External Calibration action commit EEPROM Update When an adju...

Page 10: ...he As Left status of the device Note If any test fails Reverification after performing an adjustment verify that you have met the Test Conditions before returning your device to NI Refer to World Wide Support and Services for assistance in returning the device to NI Specifications Refer to the NI PXIe 4304 4305 Specifications document for detailed PXIe 4304 4305 specification information World Wid...

Reviews: