NI PXIe-4304/4305 Calibration Procedure

Initial Setup

Refer to the 

NI PXIe-4304/4305 and TB-4304 Installation Guide and Terminal Block 

Specifications

 for information about how to install the software and hardware and how to 

configure the device in NI Measurement & Automation Explorer (MAX).

Note

When a device is configured in MAX, it is assigned a device identifier. Each 

function call uses this identifier to determine which DAQ device to verify or to verify 
and adjust. This document uses 

Dev1

 to refer to the device name. In the following 

procedures, use the device name as it appears in MAX.

Note

If you are using the PXIe-4139 as a source, no self-calibration is required.

Accuracy Verification

The following performance verification procedures describe the sequence of operations and 
provides the test points required to verify the PXIe-4304/4305. The verification procedures 
assume that adequate traceable uncertainties are available for the calibration references.

The PXIe-4304/4305 has 32 independent analog input channels. You can verify the accuracy of 
any or all of the channels depending upon your desired test coverage.

Complete the following steps to verify the voltage mode accuracy of the PXIe-4304/4305.
1.

Set the PXIe-4139 voltage output to zero volts.

2.

Connect the PXIe-4139 and PXI-4071 to the TB-4304 as shown in Figure 2.

3.

Use Table 3 to configure the PXIe-4139 to output a Test Point value for the appropriate 
range shown in Table 6, beginning with the values in the first row.

Table 3.  

PXIe-4139 Voltage Output Setup

Configuration

Value

Function

Voltage output

Sense

Remote

Range

600 mV range for test points of 0 V

60 V range for all other test points

Current Limit

20 mA

Current Limit Range

200 mA

Summary of Contents for PXIe-4304

Page 1: ... 10 Specifications 10 World Wide Support and Services 10 Software Calibrating the NI PXIe 4304 4305 requires the installation of NI DAQmx on the calibration system Driver support for calibrating the PXIe 4304 4305 was first available in NI DAQmx 15 1 1 For the list of devices supported by a specific release refer to the NI DAQmx Readme available on the version specific download page or installatio...

Page 2: ... reference information NI PXIe 4304 4305 Specifications PXIe 4304 4305 specifications and calibration interval NI DAQmx Readme Operating system and application software support in NI DAQmx NI DAQmx Help Information about creating applications that use the NI DAQmx driver LabVIEW Help LabVIEW programming concepts and reference information about NI DAQmx VIs and functions NI DAQmx C Reference Help R...

Page 3: ...able 1 Table 1 Recommended Equipment for PXIe 4304 4305 Verification and Adjustment Equipment Recommended Model Requirements DMM PXI 4071 Use a DMM that has an accuracy of 32 ppm or better when measuring voltages higher than 10 V and an accuracy of 2 1 μV or better when measuring 0 V PXI Express Chassis PXIe 1062Q If this chassis is unavailable use another PXI Express chassis such as the PXIe 1082...

Page 4: ...I PXIe 4304 4305 and TB 4304 User Guide and Terminal Block Specifications 2 Configure the PXIe 4139 to voltage output mode and enable remote sensing Connect the PXIe 4139 output to the TB 4304 as shown in Figure 2 Table 2 TB 4304 Analog Signal Names Signal Name Signal Description AI Positive input voltage terminal AI Negative input voltage terminal AIGND Analog ground input NI TB 4304 AI3 AI3 AI2 ...

Page 5: ...d pair wire to eliminate noise and thermal offsets Maintain an ambient temperature of 23 C 5 C The PXIe 4304 4305 temperature will be greater than the ambient temperature Maintain an internal device temperature range of Tcal 5 C for the PXI 4071 1 Keep relative humidity below 80 Allow a warm up time of at least 15 minutes to ensure that the PXIe 4304 4305 measurement circuitry is at a stable opera...

Page 6: ...the sequence of operations and provides the test points required to verify the PXIe 4304 4305 The verification procedures assume that adequate traceable uncertainties are available for the calibration references The PXIe 4304 4305 has 32 independent analog input channels You can verify the accuracy of any or all of the channels depending upon your desired test coverage Complete the following steps...

Page 7: ...ement Range 1 V range for test points of 0 V 100 V range for all other test points Digital Resolution 7 5 digits Aperture Time 100 ms Autozero On ADC Calibration On Input Impedance 10 MΩ DC Noise Rejection High Order Number of Averages 1 Power Line Frequency Dependent upon the local power line characteristics Table 5 AI Voltage Mode Setup Configuration Value Channel Name Dev1 aix where x refers to...

Page 8: ...B 4304 as shown in Figure 2 3 Call the DAQmx Initialize External Calibration function with the following parameters Device In Dev1 Password1 NI 4 Call the 4304 4305 instance of the DAQmx Setup SC Express Calibration function with the following parameters calhandle in calhandle output from DAQmx Initialize External Calibration rangeMax Appropriate Range Max starting with the first Test Point value ...

Page 9: ... for the remaining Test Point values from Table 7 for the corresponding range that was configured in step 4 10 Repeat steps 4 through 9 for the remaining ranges from Table 7 11 Call the 4304 4305 instance of the DAQmx Adjust SC Express Calibration function with the following parameters calhandle in calhandle output from DAQmx Initialize External Calibration action commit EEPROM Update When an adju...

Page 10: ...he As Left status of the device Note If any test fails Reverification after performing an adjustment verify that you have met the Test Conditions before returning your device to NI Refer to World Wide Support and Services for assistance in returning the device to NI Specifications Refer to the NI PXIe 4304 4305 Specifications document for detailed PXIe 4304 4305 specification information World Wid...