cRIO-9052 Operating Instructions
16
ni.com
Environmental
The cRIO-9052 is intended for indoor use only. For outdoor use, mount the
CompactRIO system in a suitably rated enclosure.
Operating temperature
(IEC 60068-2-1, IEC 60068-2-2) ...........–40 to 70 °C
Note
To meet this operating temperature range, follow the guidelines in the installation
instructions for the CompactRIO system.
Storage temperature
(IEC 60068-2-1, IEC 60068-2-2) ...........–40 to 85 °C
Ingress protection ...................................IP 40
Operating humidity
(IEC 60068-2-56) ...................................10 to 90% RH, noncondensing
Storage humidity
(IEC 60068-2-56) ...................................5 to 95% RH, noncondensing
Maximum altitude...................................2,000 m
Pollution Degree (IEC 60664) ................2
Shock and Vibration
To meet these specifications, you must panel mount the CompactRIO
system and affix ferrules to the end of the terminal wires.
Operating vibration
Random (IEC 60068-2-64)..............5 g
rms
, 10 to 500 Hz
Sinusoidal (IEC 60068-2-6) ............5 g, 10 to 500 Hz
Operating shock
(IEC 60068-2-27) ...................................30 g, 11 ms half sine
50 g, 3 ms half sine,
18 shocks at 6 orientations