4 AI, ±10 V, 16 Bit, 500 kS/s/ch Simultaneous
Page 1: ...GETTING STARTED GUIDE NI 9222 4 AI 10 V 16 Bit 500 kS s ch Simultaneous...
Page 2: ...Refer to the documentation for each component in the system to determine the safety and EMC ratings for the entire system Safety Guidelines Operate the NI 9222 only as described in this document Cauti...
Page 3: ...Measurement Category I Withstand 1000 Vrms verified by a 5 s dielectric withstand test Measurement Category I is for measurements performed on circuits not directly connected to the electrical distri...
Page 4: ...circuits are not intended for direct connection to the MAINS building installations of Measurement Categories CAT II CAT III or CAT IV Safety Guidelines for Hazardous Locations The NI 9222 is suitabl...
Page 5: ...losure rated to at least IP54 as defined by IEC EN 60079 15 Special Conditions for Hazardous Locations Use in Europe and Internationally The NI 9222 with screw terminal has been evaluated as Ex nA IIC...
Page 6: ...shall only be used in an area of not more than Pollution Degree 2 as defined in IEC 60664 1 Caution The system shall be mounted in an ATEX IECEx certified enclosure with a minimum ingress protection...
Page 7: ...and television reception and prevent unacceptable performance degradation install and use this product in strict accordance with the instructions in the product documentation Furthermore any changes o...
Page 8: ...e specified EMC performance the length of all I O cables must be no longer than 30 m 100 ft Cable Requirements for EMC Compliance Select and install cables for the NI 9222 with screw terminal in accor...
Page 9: ...for NI 9222 with screw terminal for EMC Compliance NI 9222 Shield AI0 AI0 Signal Source Shield AI1 AI1 Signal Source Shield AI2 AI2 Signal Source Shield AI3 AI3 Signal Source NI 9222 Getting Started G...
Page 10: ...certificate or look for the Lloyd s Register mark on the product Caution In order to meet the EMC requirements for marine applications install the product in a shielded enclosure with shielded and or...
Page 11: ...s Operating temperature IEC 60068 2 1 IEC 60068 2 2 40 C to 70 C Operating humidity IEC 60068 2 78 10 RH to 90 RH noncondensing Pollution Degree 2 Maximum altitude 5 000 m Indoor use only Note Refer t...
Page 12: ...NI 9222 Pinout AI0 AI0 AI1 AI1 AI2 AI2 AI3 AI3 0 1 0 1 0 1 0 1 AI0 AI0 AI1 AI1 AI2 AI2 AI3 AI3 12 ni com NI 9222 Getting Started Guide...
Page 13: ...nnection AI Negative analog input signal connection Grounded Connections NI 9222 AI AI Signal Source Make sure the voltage on the AI and AI connections are in the channel to earth safety voltage range...
Page 14: ...e sure that devices you connect to the NI 9222 are compatible with the module specifications You must use 2 wire ferrules to create a secure connection when connecting more than one wire to a single t...
Page 15: ...he NI 9222 with screw terminal Use ferrules to terminate wires to the detachable connector Use the NI 9971 backshell kit Overvoltage Protection The NI 9222 provides overvoltage protection for each cha...
Page 16: ...als RELATED INFORMATION C Series Documentation Resources ni com info cseriesdoc Services ni com services Installs with the software NI 9222 Datasheet NI RIO Help LabVIEW FPGA Help NI 9222 Datasheet NI...
Page 17: ...duct Product registration facilitates technical support and ensures that you receive important information updates from NI A Declaration of Conformity DoC is our claim of compliance with the Council o...