3. If the buoy is deployed in saltwater, inspect the sacrificial zinc anodes and replace as needed.
.
4. Inspect sensor cables and calibrate instruments per manufacturer’s recommendations.
a. Before disconnecting any sensor cables, it is recommended to remove power from the buoy by
disconnecting the UW-6 pin plug from the SOLAR port on the data logger/data well lid.
i. Make sure to protect the plug with a cap and the port with a port plug to avoid water
intrusion while disconnected.
b. Carefully disconnect and inspect any UW or MCIL connectors.
i. Follow the article below for proper connector operation and maintenance.
MCBH & MCIL Connector Operation & Maintenance
5. Be sure to reconnect the UW-6 plug in the SOLAR port when finished.
a. For systems with a NexSens data logger, an
indicates that the logger is powering
up.
b. Allow time for a measurement cycle and check readings on WQData LIVE before leaving the
site to make sure everything is in working order.
NexSens CB-ZA Sacrificial Zinc
Anode
48