PPA45xx Quick User Guide
Page 46
The resulting graph and results table with smoothing applied are displayed
below;
The displays above show how with smoothing applied, the data is smoothed out
over the resultant timescale and displaying an intermediate value for every
update window during the step between the two peak voltage values
Note: each speed parameter has its own time constant for smoothing and data
updates as shown in the table below
Speed
Update
Rate:
(speed only)
Normal
Smoothing:
applicable to
relevant
speed
Slow
Smoothing:
applicable to
relevant
speed
Very Slow
10s
48s
196s
Slow
2.5s
12s
48s
Medium
0.333s
1.5s
6s
Fast
0.05s
0.2s
0.8s
Very Fast
0.0125s
0.05s
0.2s
Window
Manually
Input speed
setting
different to 5
pre selected