Newport Oriel OPS-Q1000 User Manual Download Page 56

90074225  MOPS-Q  QTH  Lamp  Pow er Supplies 

Page  56 

 

 

19 TROUBLESHOOTING

 

 

This  chart provides  the  basic troubleshooting information for  the  Newport  Model  OPS power  supplies 
when used with Newport lamp housing.   Contact a Newport sales engineer or  your local representative if 
more information is required. 

 

Symptom

 

Action

 

Power supply does not turn on, e.g. 

  

Fan not turning 

  

Display does not turn on 

Check 

  

AC power cable is properly connected to AC INPUT 
terminal 

  

AC Main Power switch is switched to the ON position 

  

AC power at wall outlet 

Fault light 

  

Display shows 

“Interlock 

Ope

n” 

Turn off AC power and Check 

  

Lamp interconnection connection cable(s) are firmly fixed 
to connectors on power supply and lamp housing 

  

Door to lamp housing is closed 

  

If the lamp was running before this error message 
appeared and the cooling fan of the lamp housing was 
operational, this indicates too high a temperature inside 
the lamp housing. Confirm there is no blockage of the 
cooling fan or vents of the lamp housing. 

Fault light 

  

Display shows 

“Ignition 

Fail

ed” 

  

Ticking sound heard from top 
of lamp housing during 
ignition cycle approximately 
once per second 

This message is displayed usually when trying to ignite and 

ignition fails. It is a result of no current flow from supply after 5 
second of trying to ignite the lamp. 

  

Turn off the power supply and disconnect the Lamp 
Interconnection Cable(s) between the lamp housing and 
the power supply. 

  

Check lamp connection and polarity inside the lamp 
housing.  A lamp installed upside down or with reversed 
polarity should be removed and replaced immediately. 

  

Reconnect the Lamp Ignition Cable(s) between the lamp 
housing and power supply.  Turn on the power supply. 
Re-attempt lamp ignition. 

  

Confirm the hours of usage on the lamp.  A lamp 
approaching the end of its rated lifetime will be difficult to 
ignite. 

  

Attempt ignition with a new lamp. 

  

If fault repeats contact Newport for RMA information. 

Shutter icon indicates the opposite of 
its actual status with external shutter 
TTL  control. 

In the System1 setup menu, change the Shutter setting to reflect 

the required TTL  signal polarity. 

Summary of Contents for Oriel OPS-Q1000

Page 1: ...s OPS Q250 OPS Q1000 Oriel Quartz Tungsten Halogen Lamp Power Supplies User s Manual Family of Brands ILX Lightwave New Focus Ophir Corion Richardson Gratings Spectra Physics 90074225MOPS Q Rev B...

Page 2: ...NG THE LAMP ON OFF 18 4 3 LIGHT INTENSITY CONTROL 19 5 OPERATING MODES 20 5 1 POWERMODE 20 5 2 CURRENT MODE 20 5 3 INTENSITY MODE 20 5 3 1 Dosed Mode 20 6 POWER AND CURRENT OPERATION MODES 21 6 1 POWE...

Page 3: ...el Connections 13 Figure 4 The LCD display of the OPS Q Model Power Supply 17 Figure 5 USB Connectorfor FirmwareUpdates 25 Figure 6 The LCD screen image displayedduring firmwareand LCD GUI interfaceup...

Page 4: ...e as theOperatingMode of the OPS Q Power Supply 37 Figure 16 Idle FunctionSetup 39 Figure 17 Shutter Status Indicator on Display 40 Figure 18 Shutter Status Symbols 40 Figure 19 Manual Shutter Mode Sh...

Page 5: ...mishandling could result in irreparable damage to the equipment and personal injury or death EUROPEAN UNION CE MARK The presence of the CE Mark on Newport Corporation equipment means that it has been...

Page 6: ...ower sources before servicingthe equipment To avoid explosion do not operate this equipment in an explosiveatmosphere Qualifiedservice personnel should perform safety checks after any service If this...

Page 7: ...wer supply A substitute line cord may not be rated for high current 1 5 LAMP HANDLING Read all informationand warnings provided with lamp Never touch any lamp or the reflector s inner surface with bar...

Page 8: ...nel for easy navigation of power supply features LAMP and SHUTTER buttons located on front panel for manual control of lamp ignition and external shutter Icons on the LCD display show the ignition and...

Page 9: ...90074225 MOPS Q QTH Lamp Pow er Supplies Page 9 2 2 DIMENSIONS Figure 1 Dimensionaldrawing of the OPS SeriesPower Supplies...

Page 10: ...881 66882 66883 66884 67009 67011 60043 n a 6337 QTH 50W LONG FILAMENT 60000 Housing with 60090 Interface Kit 60043 66881 66882 66883 66884 67009 67011 60043 n a 6333 QTH 100W 60000 Housing with 60090...

Page 11: ...e date When handling this power supply hold it from underneath the housing Never use any knobs buttons or cables to carry or locate the power supply WARNING Do not attempt tooperatethis equipment if t...

Page 12: ...etting and monitoring various features of the power supply a Lamp OperatingMode b Manual TimedShutter Control DisplayFeatures c Saving loadingparticularsetups C SHUTTER This button is for manual contr...

Page 13: ...o the main circuitry of the power supply There is no output from the power supply until the manual or external command to supply power to the lamp is received with the interlock conditionsatisfied 3 4...

Page 14: ...this system Never open the OPS cover and attempt to work inside Use the power cord provided with the power supply Never alter the power cord in any way Be sure to periodically inspect the power cord a...

Page 15: ...nectioncable provided by a third party vendor or made on the user s end could damage the power supply and lamp housingand pose an electrical shock risk to the user The connectorpin assignmentsare list...

Page 16: ...L signal may be input to the TRIGGER INPUT connectoror a remote switch closure maybe wired to the connector G SHUTTER OUTPUT The BNC connector labeled SHUTTER OUTPUT outputs a TTL signal to an externa...

Page 17: ...the power supply and related components lamp lamphousing cable s are preparedfor lampignition or displays Interlock Open indicating a problem with the system setup Similar error messages such as Cool...

Page 18: ...the vertical menu Press the vertical menu button next to Pset Iset and use the knob to set this parameter when a red bullet appears next to the current value Then set Pmax Imax This is typically set...

Page 19: ...d this value This value is to be used as the Flux Set set point iii If the recorded value shows value of 2 5uA 25uA or 250uA or 2500uA this indicates that the flux reading has overflowed with existing...

Page 20: ...000 Q Series and 45 1000 W Oriel Research Lamp Housings adapters included and OPS model power supply While operating in Intensity Mode varying the Flux Set parameter allows the user to increase decrea...

Page 21: ...djust the Pset and Pmax press the vertical menu button next to each parameter A red bullet in the box indicating that particular parameter can be changed will appear That parameter can now be changed...

Page 22: ...er supply is being operated in Power Mode this indicates that the Imax setting in Current Mode is set too low for the lamp to reach the desired Pset If this error messageappears power down the lamp Th...

Page 23: ...3 To adjust the Iset and Imax press the vertical menu button next to each parameter A red bullet in the box indicating that particular parameter can be changed will appear That parameter can now be ch...

Page 24: ...s option can be changed by pressing the vertical menu key and using the external knob to select On or Off when a red bullet appears to the left of the current Audio setting LampType This is a read onl...

Page 25: ...e or reversion if a previous version exists on the USB flash drive will occur The LCD GUI image file naming format will be imNN_XXX hex with NN being the image number 0 5 and XXX being the version num...

Page 26: ...by the procedureexplainedpreviously in this section of the manual 8 INTERLOCK CABLES CONNECTIONS TO LAMP HOUSINGS A lamp interconnection cable must be purchased for use with a Research Lamp Housing b...

Page 27: ...the previous page are used The use of a lamp housing interconnectioncableprovided by a third party vendor or made on the user s end could damage the power supply and lamp housing and pose an electrica...

Page 28: ...esearch Lamp Housingfor use with an OPS Q Model Power Supply 8 3 SAFETY INTERLOCK Newport s Oriel Power Supplies have a safety interlock feature which must be satisfied before the power supply will po...

Page 29: ...e to factors beyond the control of the power supply such as lamp aging ambient temperature fluctuations and electrode erosion For applications in which especially high light output stability is requir...

Page 30: ...active area of the detector The predominant cause of dark current is thermionic emissions in which the detector outputs a current value as a result of ambient temperature To counteract this silicon ph...

Page 31: ...ears the CoolTemp setting needs to be adjusted Sometimes external conditions such as an ambient operating temperature higher than typical room temperature 21 3 C can prevent the OPS Q Power Supply fro...

Page 32: ...creases the frequency at which the photosensor must be returned for re calibration The OPS Q Power Supply is also specifically designed to control and maintain the operating temperature of the 71582 t...

Page 33: ...couple the 71582 detector head to the Series Q Housing If an Oriel Products Research Lamp Housing is being used use the 68952 Adapter Kit to couple the 71582 detector head to the Research Lamp Housin...

Page 34: ...n the rear of the 71582 Detector Head Then connect the 9 pin input connector on the other end of the 70062 to the 9 pin connectorheadon the rear panel of the OPS QPower Supply under thelabel markedTE...

Page 35: ...S Q Power Supply Figure 16 The 70062 left and 70018 right requiredfor interfacingthe 71582 DetectorHead with OPS Q QTH Lamp PowerSupply 3 Ignite the lamp using either the Power or Current Operation Mo...

Page 36: ...der the Setup indicator on the LCD display until System 2 appears with the correspondingvertical menu shown above Then press the vertical menu button to the right of the AutoSet indicator This automat...

Page 37: ...supplied to the lamp in order to decrease the lamp s power output accordingly and increasing the Flux Set value will increase the current power supplied to the lamp in order to increase the lamp s pow...

Page 38: ...stems where the 71582 TE Cooled Si detector is not coupled to the lamp housing and located after the shutter relative the light path In this type of setup the 71852 is not exposed to any light when th...

Page 39: ...es are set by the user ExternalTTL control The shutter opens or closes based on an external TTL low or high signal ExternalTTL trigger The shutter begins a preset time or dose routine when triggered b...

Page 40: ...will open or close based on the input of a TTL high or low signal from the Trigger Input connector on the back of the power supply If the shutter status symbol on the power supply display indicates th...

Page 41: ...he button labeled SHUTTER on the front face of the power supply Figure 22 ManualShutter Button 11 3 TTL SHUTTERCONNECTION Any shutter that can be driven by a TTL signal may be controlled by the OPS po...

Page 42: ...System 1 setup menu as shown below Press the vertical menu button next to the Shutter indicator When a red bullet appears next to the current Shutter setting use the knob to adjust to Pos or Neg as re...

Page 43: ...lamp is being operated in Power Current and Intensity Modes 1 To operate the shutter in Timed Shutter Mode press the horizontal menu key under the Shutter indicator until Timed appears under the Shut...

Page 44: ...next to the Back indicator to bring back the first menu shown on the previous page When the shutter is operated in Timed Exposure Mode new display options are available Pressing the horizontal menu k...

Page 45: ...t the Flux Set Flux Multiplier Iidle and Imax parametersas necessary required 2 Press the horizontal menu key under the Shutter indicator until Dosed appears in the indicator with the correspondingver...

Page 46: ...D screen shows that Dose and Total Dose are now available display options Total Dose is the accumlated Dose the lamp has been allowed to irradiate light onto the irradiationarea over multiple cycles 4...

Page 47: ...gger Input connector while the cycle is in progress If the shutter is set to Manual mode and a TTL signal is sent to the Trigger Input connector the shutter remains open for the duration of the Low si...

Page 48: ...nterlock Open Lamp housing connectioncable became loose or lamp housing door were opened when the lamp was ignited Ignition Failed Ignition failure during operation of the lamp A very low current draw...

Page 49: ...e settings are automatically loaded Because of the versatility of the OPS power supplies Oriel recognizes that this unit may be utilized in a wide variety of projects Therefore the power supply was de...

Page 50: ...ext to that Setup space The settings the power supply is currently operating under are now saved under the selected Setup space The following parameters are saved when a Setup is saved Lamp OperatingM...

Page 51: ...and or damage to property Figure 28 LoadingSaved SettingsFrom Memory Loading a previously saved Setup is very similar to the methodused to save that Setup Press the horizontal menu button under the S...

Page 52: ...gradually reducing the light output With tungsten halogen lamps the halogen gas effectively removes the deposited tungsten and returns it to the hot filament leaving the inside of the envelope clean a...

Page 53: ...anel Use the scroll wheel on the power supply until Lamp Hours is selected Press the Display or CLEAR button in order to view the number of hours the lamp has been turned on Please note that Lamp Hour...

Page 54: ...t ignited The Reset Lamp Hours feature was integrated into the power supply to allow a user to monitor lamp lifetime from installation or its first usage If necessary the monitoring of any time interv...

Page 55: ...eter Resolution Current 0 01 A Safety Interlock Voltage 12 VDC GND OperatingMode Constant Current Power or Intensity OperatingTemperature 10 45 C Max Relative Humidity 80 non condensing Max Weight lb...

Page 56: ...the cooling fan or vents of the lamp housing Fault light Display shows Ignition Failed Ticking sound heard from top of lamp housing during ignition cycle approximately once per second This messageis d...

Page 57: ...supply If fault repeats contact Newport for RMA information Fault light Display shows Power SupplyFailed This messageis displayed when the power supplycannot be enabled for the lamp or there is a shor...

Page 58: ...lux reading of the LIK LMP s 71582 detector head out of the recordablerange of the power supply 1 If the 71582 detector head is directlyconnectedto lamp housingswitch to the other Disk with Aperture i...

Page 59: ...resthat the product ProductName Optical Power Supply Model Numbers OPS Q250 OPS Q1000 Type of equipment Electrical equipmentfor measurement control and laboratory use in industrial locations conformst...

Page 60: ...used Pin 8 Not used Pin 9 Not used Figure 30 RS232 Connector Serial Port Setup 9600 Baud 8 bits 1 Stop No Parity No Flow Control Syntax ASCII Text Received commandlines are to be terminatedby CR Tran...

Page 61: ...erlock isopen ESR none ESRXX HEX Bit5 Command Error Bit0 Operation Complete A responseof 64 indicatesthattheunit did notunderstandthe command Aresponse of 1 indicatesthatthe prior command has been pro...

Page 62: ...mand for Hex valuedefinition Reset power supply to factory default RSTHRS none ESRXX SeeESRcommand for Hex valuedefinition Resetlamp hoursto 0 MODE i i 0 Power mode i 1 Currentmode ESRXX SeeESRcommand...

Page 63: ...imit when lamp isoff P LIM iiii ESRXX SeeESRcommand for Hex valuedefinition Set power limit when lamp isoff STARTEXP none 0 for success Starts exposurer running STOPEXP none 0 for success Stops exposu...

Page 64: ...Queries thetotal accumulatesopen exposuretime OPENTIME none hh mm ss th Queries theopen timefor exposures cycles OPENTIME hh mm ss th 0 for success Sets theopen timefor exposures cycles EXPOPEN none h...

Page 65: ...ueries thefluxvalue FLUXMULT none ffff ff Queries thefluxmultiplier FLUXMULT ffff ff 0 for success Sets theflux multiplier FLUXCONTROL none 1 on 0 off Queries statusonflux control DETECTOR none ffff f...

Page 66: ...to find range bit3 auto setup with cooler notready bit4 auto setup failed to seesignal bit5 auto setup detector saturated bit6 auto setup failed to find setpoint bit7 received command notunderstood b...

Page 67: ...ices worldwide Every year the Newport Resource catalog is hailed as the premier sourcebook for those in need of advanced technology products and services It is available by mail request or through New...

Page 68: ...RMA number must be obtained prior to shipment to Oriel Instruments This RMA number must appear on both the shipping containerand the packagedocuments Return the product to Oriel Instruments freight pr...

Page 69: ...ed at Buyer s expense to Oriel facility provided that Buyer notifies Oriel in writing promptly after discovery of the defect or nonconformity and within the Warranty Period Products may only be return...

Page 70: ...trade secrets of Newport and are for use only in the manner expressly permitted Newport claimsand reservesall rightsand benefitsafforded under law in the Programsprovided by Newport Corporation Newpor...

Reviews: