45
Details of data package format:
Byte_0 (status byte)
0x00 = regular energy value
0x01 = energy value OVER (out of range)
0x0A = frequency value
Byte_1
Timestamp byte 0 (LSB)
Byte_2
Timestamp byte 1
Byte_3
Timestamp byte 2 (MSB)
Byte_4 (FLOAT)
Value (energy or frequency) byte 0 (Least significant)
Byte_5 (FLOAT)
Value (energy or frequency) byte 1
Byte_6 (FLOAT)
Value (energy or frequency) byte 2
Byte_7 (FLOAT)
Value (energy or frequency) byte 3 (Most significant)
4. OTHER MODES:
At present the CS modes defined are as described above. In addition there are other options that could be added in
the future according to customer requirements. Please be in contact with Ophir via your local sales agent or directly
to the R&D Department (contact details below in Appendix 3) if you need assistance or if you have a request for a
special measurement mode that is not currently supported.