7
EDH0351En1012 — 10/18
ONE-XY Series Integrated XY Linear Motor Stages
WARNING
GROUNDING: The stage's protective ground is located on pin #8 of the
motor power connector.
If you are using grounds other than those provided by Newport, you
must connect pin #8 to a ground connection.
4.5.2
Encoder and End-of-Runs (SUB-D15M Connector)
4.6
ONE-XY Cable Wiring
All ONE-XY stages are delivered with the three cables required for
operation. The wiring diagrams and connectors for these cables are
provided in the following diagrams. When operating the non-Newport
controllers, it is recommended to adhere to the wiring conventions
presented here and to use cabling with similar characteristics.
NOTE
This is the wiring of the cables that provided with the connection to
Newport Controllers, for third party controllers, we recommend using a
cable with similar characteristics.
8
15
9
1
1
Encoder Phase B
2
Shield Ground
3
Encoder Phase A
4
+5 V Encoder
5
N.C.
6
– End-of-Run
7
Index Pulse /I
8
+ End-of-Run
9
Encoder Phase /B
10
Shield Ground
11
Encoder Phase /A
12
+5 V
13
N.C.
14
Index Pulse I
15
N.C.
M
o
to
r
E
n
c
o
d
e
r
&
E
n
d
-o
f-
R
u
n
s
SUB-D9M
Connector
1
2
3
4
6
7
8
Connector
Cap
ONE-XY
Stage
Phase U Motor
Phase U Motor
Phase V Motor
Phase V Motor
Phase W Motor
Phase W Motor
Ground
Connector
Cap
Sub-D15F
Connector
Encoder Phase A
Encoder Phase /A
Encoder Phase B
Encoder Phase /B
Index Pulse I
Index Pulse /I
+ End-of-Run
- End-of-Run
+5 V
Encoder
+5 V
N.C.
N.C.
N.C.
Shield Ground
Connector
Cap
SUB-D15M
Connector
3
11
1
9
14
7
8
6
4
12
5
13
15
2-10
Connector
Cap