![Newport INFCP Quick Start Download Page 1](http://html1.mh-extra.com/html/newport/infcp/infcp_quick-start_1680038001.webp)
INFCP and INFCP-xxxB
Process Meter
Using This Quick Start Manual
Use this Quick Start Manual to set up your Process Meter and begin
operation. Information is provided on how to:
•Connect ac power
•Set basic options for operation
•Connect the sensor
•Scale the meter.
Features with are for the “B” version which has three-color
programmable “Big” LED display - All segment characters shown are
for the “B” version.
IMPORTANT:
For complete information on all setup options,
please refer to the Operator’s Manual.
Mount the Unit
1. Cut a panel opening using the
dimensions shown to the right.
2. Position the unit in the opening,
making sure the front bezel is
flush with the panel.
3. Install retaining clips on both
sides of the meter and tighten
against the panel.
Wiring
1. Remove the panel at the back of the unit.
2. Locate the TB1 connector.
3. Insert the correct wire in each terminal as shown in the following
figure and tighten the lockdown screws.
4. Tug gently on the wires to verify the connections.
Connect the Sensor
1. Locate the TB2 connector on the rear of the unit.
2. Attach the sensor wires and tighten the lockdown screws.
The diagram below shows the wiring for 4–20 mA sensors
with internal excitation.
3. Tug gently on the wires to verify the connections.
4. Replace the panel at the back of the unit.
NOTE:
Refer to the Operator's Manual for setup requirements
for other sensor types.
Current Input Connections (4–20 mA)
with Internal Excitation
Using the Configuration Menu
To configure the meter, you use the buttons on the front panel.
To:Take This Action:
Display the
Press the
MENU
button. The first function
Configuration Menu on the menu,
INPT
, displays.
Select a submenu
1. Press
MENU
until the function you
function
want is shown.
2. Press
䊳
/TARE
.
The information you can change flashes.
Select a value
1. Press
䊱
/MAX
to display the option
for that submenu
you want.
function
2. Press
MENU
to store it.
STRD
quickly flashes, indicating that
the selection has been stored in memory.
Then the next menu function displays.
Go back to previous Press
RESET
once.
menu function
Exit the
Press
RESET
twice. The unit displays
Configuration
RST
as it reinitializes. When a numeric
Menu
value displays, the unit is in run mode.
(Optionally, you can press
MENU
to
move through all the menu functions
until the unit reinitializes.)
NOTE: This Quick Start Manual includes specific configuration
parameters for transducers with an output range of 4–20 mA and
24 V excitation. Other sensor types may require different
parameters or additional ones. When this is the case, we refer you
to the Operator's Manual for detailed instructions.
2
3
4
Warning: Do not connect AC power to your device until you
have completed all input and output connections. This
device must only be installed by a specially trained
electrician with corresponding qualifications. Failure to
follow all instructions and warnings may result in injury!
AC Powered Unit Connections
DC Powered Unit Connections
To Set the Input Type
1. Press
MENU
until the unit displays:
2. Press
䊳
/TARE
. The unit displays:
3. For this application you want
0-20
. If
0-20
is not
displayed, press
䊱
/MAX
until it appears. Other choices
are
100M
,
50M
,
10V
and
5V
.
NOTE: Refer to the Operator's Manual for more information
on changing ranges.
4. Press
MENU
to select the sensor shown. The meter
displays the next menu item. If you changed input type,
the meter displays:
To Set the Decimal Point
1. If it's not already shown, press
MENU
until the unit displays:
2. Press
䊳
/TARE
. The unit displays:
3. Press
䊱
/MAX
to move the decimal
point to the desired location. The
choices are
FFFF
,
F.FFF
,
FF.FF
,
and
FFF.F
.
4. Press
MENU
to select the decimal point position shown.
The unit displays:
MENU
SUBMENU
䊳
/TARE
DESCRIPTION
INPT
100M
,
±50M
,
!10V
,
!±5V
,
0-20
* Input
DEC.P
FFFF
*,
F.FFF
,
FF.FF
,
FFF.F
Decimal Point
RD.S.O
IN!1
,
RD!1
,
IN!2
,
RD!2
Scale and Offset
RD.CF
R.1=T
*,
R.1=N
Reading Configuration
R.2=0
,
R.2=1
,
R.2=2
,
R.2=3
,
R.2=4
*
R.3=F
*,
R.3=U
COLR
GRN
,
RED
,
AMBR
Display Color
S1.CF
S.1=A
*,
S.1=B
Setpoint 1 Configuration
S.2=U
*,
S.2=L
S2.CF
S.1=A
*,
S.1=B
Setpoint 2 Configuration
S.2=U
*,
S.2=L
S1.DB
0003
*
Setpoint 1, Deadband
S2.DB
0003
*
Setpoint 2, Deadband
OT.CF
O.1=E
*,
0.1=D
Analog Output
O.2=C
*,
0.2=V
Configuration
O.3=A
*,
0.3=P
O.4=D
,
0.4=R
O.5=F
,
0.5=H
P.BND
0000
shown if 0.3 = P
Proportional Band
M.RST
0000
shown if 0.3 = P
Manual Reset
OT.S.O
RD!1
,
OUT1
,
RD!2
,
OUT2
Output Scale & Offset
LK.CF
RS=E
*,
RS=D
Lockout Configuration
SP=E
*,
SP=D
L3=0
*,
L3=1
BRIT
M.BrT
,
L.BrT
,
H.BrT
Display Brightness
External Fuse Required:
Time-delay, UL 248-14 listed
Time-lag, IEC 127-3 recognized
175 mA (115 Vac line)
125 mA (115 Vac line)
80 mA (230 Vac line)
63 mA (230 Vac line)
10-32 Vdc, 7.5 W
DEVICE
* Factory Default Settings
When using DC power, do not use internal excitation or
Isolated Analog Output for high color brightness. For
low or medium brightness, internal excitation is limited
to 24 V @ 25 mA, 5 V, 10 V, 12 V @ 35 mA.
In order to maintain the same degree of protection as the
AC units, always use a Safety Agency Approval DC source
with the same Overvoltage Category and Pollution Degree.
For immediate technical or application assistance please call:
Newport Electronics, Inc.
2229 South Yale Street • Santa Ana, CA • 92704 • U.S.A.
TEL: (714) 540-4914 • FAX: (714) 546-3022
Toll Free: 1-800-639-7678 • http://www.newportUS.com • e-mail:[email protected]
ISO 9001 Certified
Newport Technologies, Inc.
976 Bergar • Laval (Quebec) • H7L 5A1 • Canada
TEL: (514) 335-3183 • FAX: (514) 856-6886
Toll Free: 1-800-639-7678 • http://www.newport.ca • e-mail:[email protected]
Newport Electronics, Ltd.
One Omega Drive • River Bend Technology Centre
Northbank, Irlam • Manchester M44 5BD • United Kingdom
Tel: +44 161 777 6611 • FAX: +44 161 777 6622
Toll Free: 0800 488 488 • http://www.newportuk.co.uk • e-mail:[email protected]
Newport Electronics B.V.
Postbus 8034 • 1180 LA Amstelveen • The Netherlands
TEL: +31 20 3472121 • FAX: +31 20 6434643
Toll Free: 0800 0993344 • http://www.newport.nl • e-mail: [email protected]
Newport Electronics spol s.r.o.
Rudé armády 1868, 733 01 Karviná 8 • Czech Republic
TEL: +420 69 6311899 • FAX: +420 69 6311114
Toll Free: 0800-1-66342 • http://www.newport.cz • e-mail: [email protected]
Newport Electronics GmbH
Daimlerstrasse 26 • D-75392 Deckenpfronn • Germany
TEL: 49 7056 9398-0 • FAX: 49 7056 9398-29
Toll Free: 0800 / 6397678 • http://www.newport.de • e-mail: [email protected]
Newport Electronique S.A.R.L.
9, rue Denis Papin • 78190 Trappes • France
TEL: +33 130 621 400 • FAX: +33 130 699 120
Toll Free: 0800-4-06342 • http://www.newport.fr • e-mail: [email protected]
Mexico and Latin America
TEL: 001-800-826-6342 • FAX: 001 (203) 359-7807
En Español: 001 (203) 359-7803
®
Safety Consideration
This device is marked with the international Caution symbol.
The instrument is a panel mount device protected in accordance with
Class I of EN61010-1 (AC units) and Class III (DC units). Remember
that the unit has no power-on switch. Building installation should
include a switch or circuit-breaker that must be compliant to IEC 947-1
and 947-3.
SAFETY:
•Do not exceed voltage rating on the label located on the
top of the instrument housing.
•Always disconnect power before changing signal and
power connections.
•Do not use this instrument on a work bench without its
case for safety reasons.
•Do not operate this instrument in flammable or explosive
atmospheres.
•Do not expose this instrument to rain or moisture.
EMC:
•Whenever EMC is an issue, always use shielded cables.
•Never run signal and power wires in the same conduit.
•Use signal wire connections with twisted-pair cables.
•Install Ferrite Bead(s) on signal wire close to the instrument if
EMC problems persist.