SECTION 3
OPRHT50NA B0908
3-29
NOTE:
HOME CAREGIVERS: The unit of measure, along with
other parameter settings, should be established by the patient’s
physician or Homecare Dealer.
Set Up
The Set Up parameter allows the operator to set the ventilator’s
start up settings. There are two selections, DEFAULT or USER.
Use the
▲
Up/
▼
Down button to toggle between selections.
When USER is selected, all of the ventilation parameters in effect
at shutdown will be saved. The saved parameters will appear
next time the ventilator is powered on.
When DEFAULT is selected, factory set default parameters will
appear next time the ventilator is powered on. Default
parameters are listed on pg 2-6.
NOTE:
HOME CAREGIVERS: This parameter should always be set
to USER for home use to ensure that when the HT50 is powered off
and on the physician directed ventilation settings remain in place.
Auto Panel Lock
This setting allows the user to enable (turn on) or disable (turn
off) the Auto Panel Lock feature. For a complete description of
Auto Panel Lock see pg 3-13. The Auto Panel Lock feature is
factory set at Disabled or off. Use the
▲
Up/
▼
Down buttons to
toggle between Enabled and Disabled.
NOTE:
HOME CAREGIVERS: N
EWPORT
recommends that the Auto
Panel Lock feature be enabled in homecare environments as an
added safety feature to prevent accidental changes to panel controls.
Exiting User Set Up
Exit by pressing the Silence/Reset button at any time.