Configuring The Meter
4
4.3.1 Scaling with Known Loads (On-Line Calibration) (Continued)
To scale with known inputs: apply known loads to a transducer connected to a meter, or
simulate the transducer output with a voltage or current simulator. To scale with known
inputs, follow these steps:
1.
Apply a known load equal to approximately 0% of the transducer range.
2.
Press the
MENU
button until the meter shows
RD.S.O
.
3.
Press the
∂
TARE
button. The meter shows
IN1
(Input 1).
IN1
(Input 1) is the unscaled display reading at minimum input.
4.
Press the
∂
TARE
button again. The meter shows last stored value for Input 1.
5.
Press the
∂
TARE
button once more. The meter shows the actual signal being
received.
6.
Press the
MENU
button to store this value as
IN1
(Input 1). The meter shows
RD1
(Read 1).
RD1
(Read 1) is the desired display reading at Input 1.
7.
Press the
∂
TARE
button. The meter shows the last stored value for Read 1.
8.
Press the
ß
MAX
button to change the value of your digits.
9.
Press the
∂
TARE
button to scroll horizontally to the next digit.
10.
Press the
MENU
button to store value as
RD1
. The meter shows
IN!2
(Input 2).
IN!2
(Input 2) is the unscaled display reading at maximum input.