(Figure9.1)
(Figure9.2)
Make sure that the measurement mode on the screen is
body temperature mode.
Aim the probe of the thermometer at the measured body
with a distance of 3-5cm. (Figure 9.1). Press the
measurement key, a “Bi” will sound in about 1 second,
measurement value displays on the screen (Figure 9.2),
then the measurement completes. If the measured value
exceeds the temperature alarm point (the default value is
38 ° C), five Buzzers will sound.
Make sure that the measurement mode on the screen is
material temperature mode.
Click "Setting" to switch to the surface temperature
measurement mode, aim the probe of the thermometer at
the object, press the measurement key, and the measured
temperature will be displayed in about 1 seconds.
Notice:
1.Please keep the sensor and probe cavity clean, before
and after use.
2.Please use the thermometer in a stable temperature
environment. When the ambient temperature fluctuates
greatly (like from indoor to outdoor), please leave it for
about ten minutes to measure.
3.Do not start measuring body temperature immediately
after measuring the temperature of an extremely high or
low temperature object. Please leave it for ten minutes to
measure.
4.When the measurement object comes from a place with
a large temperature difference from the measurement
environment, it should stay in the test environment for at
least five minutes.
5.Try not to measure the temperature while your forehead
is blown, drenched, sweating, applying cosmetics, etc. Do
not take temperature within 30 minutes after exercise,
bathing, and meals.
1)Body Mode
2)Surface Mode
8.Methods of measurement
10
09
3~5cm
Body