Installation
15
The laser head may now be mounted on the microscope. Tighten the set
screws which secure the base of the laser head to the laser adapter ring
(see Fig. 2-2). The red emission LED should face the operator when
standing in front of the microscope.
DANGER High Voltage and Current. The laser head and power
supply contain electrical circuits operating at lethal voltage and
current levels. Do not remove the cover from the power supply or
laser head. Serious injury or death may result.
A power line fuse is installed on the back panel of the power supply near
the power cord. A 1 amp slow blow fuse is installed. The following
model number fuses or their equivalents may be used:
Manufacturer
100-250VAC
Littlefuse
216.001 250 V/1HA
Schurter
SPT001-2504 250V/1A
Laser Head
Mounting
POWER
SUPPLY
ELECTRICAL
Power Line Fuse