19
3. Operation
Functions and Settings: Data Logging
(1) Press the MENU button to access the MENU
screen.
(2) Press the
▲
or
▼
button to select “Data logging”.
Press the
button to confirm the selection.
Menu screen
(3) The LOGGER screen will be displayed. Press the
▲
or
▼
button to select the
desired item. Press the
button to confirm the selection.
Item
Description
Start
Start data logging (page 20)
Stop
(not selectable)
Stop data logging (page 21)
Clear
Delete log data (page 22)
Interval
Set the interval between logs
(page 23)
Logger screen
To read logs, a personal commuter (page 1) and an XP302ML data logger
kit (optional) are necessary. All logs are CSV files which can be opened by
spreadsheet software (e.g. MS® Excel).
NOTE
WARNING
Do not connect/disconnect the USB cable to/from the detector in a
hazardous area.
When using the product for the first time
Format to clear all data logs (page 22).
Data logging is not available when flashing “00.01.01L00:00” is displayed
or the clock is stopped. Set the clock (page 25).
NOTE
18
3. Operation
Functions and Settings: Alarm Test
(1) Press the MENU button to access the MENU
screen.
(2) “Alarm test” is selected. Press the
button to
confirm the selection.
Menu screen
(3) The ALARM TEST screen is displayed. The light
and audio alarm start flashing and chirping.
Alarm test screen
When the alarm unit is connected
Press the MENU button to select “External AL
volume”. Alarm sound will be produced only by the
alarm unit.
Alarm test screen when
alarm unit is connected
(4) Press the ▲ or ▼ button to adjust the alarm volume.
(5) Press the
button to end the alarm test. The screen will return to the gas concentrations
screen.
If the mute mode is active, the unit
will prompt “Mute mode is active. Exit
Mute mode?” To continue the alarm
test, select “Yes” by pressing the
button. The light and audio alarm
start flashing and chirping.
NOTE
Volume adjustments made here are applied to audio gas alarm.
WARNING