40
M1288-0070
U360 Ultra-Low Temperature Freezer
User’s Manual
5.3.2 Battery
Replacement
The 6.0v Cyclon battery is mounted within the compressor housing, near the
interface board.
This is located (a) behind the lefthand base cover on the upright freezers, or (b)
in the compressor housing on the chest freezers.
WARNING!
Use only a replacement battery of the correct type and proper part
number. The battery must be fitted so the terminals correspond to
the polarity labels on the electrical panel.
To replace the battery:
1.
Turn off (
O
)
the power switch and disconnect the power supply.
2.
Remove the side cover and the screws that secure the battery to the housing
panel.
3.
Disconnect the battery terminals.
4.
Be certain, when connecting the new battery, to respect the correct polarity
(red is + positive and black is – negative).
5.
Reinstall the battery fixing screws and the side cover.
6.
Reconnect the freezer to the power supply and turn the power switch on (
I
).
5.3.3 Fuses
The following are the specifications for the fuses required by the freezers:
Main Power Supply Plug
(UK only)
13A Bussmann 1 x ¼"
†Vacuum Vent Heater
FS1 - 800mA 20x5mm Quick Acting (F) 250V Ceramic
†Main Voltage Input
FS2 - 1A 20x5mm Anti-Surge (T) 250V Ceramic
†Voltage Stabilizer Module
(option)
FS1* – 1A 20x5mm Quick Acting (F) 250V Ceramic
† These fuses must be replaced by an NBS-approved service engineer.
* This is not the same as the Vacuum Vent Heater fuse; the “FS1” designation
for both fuses refers to separate manufacturing drawings