![background image](http://html1.mh-extra.com/html/neuropace/pdms-4340/pdms-4340_user-programming-manual_1676339070.webp)
Tablet Instructions
66
RNS
®
System Programming Manual
Steps to Download a Detection Set to the Tablet
To download a detection set and save it locally to the tablet for offline use, follow these steps:
Figure 47 Start screen
1. Select the
P
ATIENT
L
IST
tile on the tablet Start screen. The patient list opens.
2. Select a patient from the patient list. The
P
ATIENT
H
OME
screen opens. Note this is different than
the tablet
H
OME
screen (shown in
on page 18).
Figure 48 Patient Home screen
3. Select the Neurostim Settings tile. The
N
EUROSTIM
S
ETTINGS
>
P
ATTER
D
ETECTION
screen opens.