NeuMoDx 900201 Operator'S Manual Download Page 2

 

NeuMoDx™ 

HCV External Controls 

INSTRUCTIONS FOR USE

  



900201

 

 

 

 

 

NeuMoDx Molecular, Inc.  

 CONFIDENTIAL

  

P/N 40600196_Rev B 

 

 

March 2019

 

  

Pg. 2 of 4 

For Ex-US Distribution Only

 

 

NeuMoDx™ 

Reagents and Consumables Required But Not Provided 

 

 

Other Equipment and Materials Required But Not Provided

 

NeuMoDx

 288 Molecular System

 [REF 500100] or 

NeuMoDx™ 96 Molecular System

 [REF 500200] 

 
 
 

WARNINGS & PRECAUTIONS    

 

 

The NeuMoDx HCV External Controls are for 

in vitro 

diagnostic use only with the NeuMoDx HCV Quant Test Strip as implemented on 

the NeuMoDx System.

 

 

 

Do not use the NeuMoDx HCV External Controls after the listed expiration date. 

 

Do not use the NeuMoDx HCV External Controls if the packaging is damaged or the kit is not frozen upon arrival. 

 

Because the NeuMoDx HCV positive controls contain intact HCV positive material, all components of the NeuMoDx HCV External 
Controls should be treated as potentially biohazardous. 

 

Always handle specimens as if they are infectious and in accordance with safe laboratory procedures such as those described in  

        Biosafety in Microbiological and Biomedical Laboratories 

1

 and in CLSI Document M29-A3.

 

Do not pipette by mouth. Do not smoke, drink, or eat in areas where specimens or reagents are being handled. 

 

Dispose of unused reagents and waste in accordance with country, federal, provincial, state and local regulations. 

 

Clean, powder-free, nitrile gloves should be worn when handling all NeuMoDx reagents and consumables.   

 

Wash hands thoroughly after performing the test. 

 

Safety Data Sheets (SDS) are provided for each reagent.  

 

 

  PRODUCT STORAGE, HANDLING & STABILITY 

 

 

The NeuMoDx HCV External Controls are shipped with dry ice to maintain a frozen state; do not use if kit contents are not frozen upon 
receipt.    

 

It is recommended that the NeuMoDx HCV External Controls

 

be stored at 

 -80°C to ensure stability.  

 

Control vials are intended for single use only. Thawed external controls may be stored at 4°C for no longer than 24 hours.   

 

Refreezing after a first thaw is not recommended.  

 

Discard any unused material after use as biohazard waste. 

 

Discard any controls that appear cloudy or contain large precipitates after thawing. 

 
  
 
 

REF 

Contents 

300300 

NeuMoDx™ 

HCV Quant Test Strip 

Dried

 

PCR reagents containing HCV-specific TaqMan

®

 probes and primers along with SPC2-specific TaqMan® probe and primers.

 

100200 

NeuMoDx™ Extraction Plate

 

Dried

 

magnetic affinity microspheres, Lytic enzyme, and sample process controls 

800200 

NeuMoDx™ 

HCV Calibrators 

Single use sets of HCV High and Low Calibrators to establish validity of standard curve

 

400600 

NeuMoDx™ 

Lysis Buffer 3

 

400100 

NeuMoDx™ 

Wash Reagent  

400200 

NeuMoDx™ 

Release Reagent  

100100 

NeuMoDx

™ 

Cartridge

 

235903 

Hamilton CO-RE Tips (300 µL) with Filters

 (available from NeuMoDx or Hamilton)

 

235905 

Hamilton CO-RE Tips (1000 µL) with Filters 

(available from NeuMoDx or Hamilton)

 